{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84832",
      "recalling_firm": "Beckman Coulter Inc",
      "address_1": "2470 Faraday Ave",
      "address_2": "N/A",
      "postal_code": "92010-7224",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Alabama,  Alaska,  Arizona,  Arkansas,  California,  Colorado,  Connecticut,  Delaware,  Florida,  Georgia,  Hawaii,  Idaho,  Illinois,  Indiana,  Iowa,  Kansas,  Kentucky,  Louisiana,  Maine,Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming, Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas,Utah,Virginia,Washington,West Virginia, Wisconsin,Wyoming      Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Curacao, Czech Republic, Egypt, Eritrea, Finland,  France, Germany, Ghana, Greece, Guatemala, Guyana, Hong Kong, India, Ireland, Italy, Japan, Kuwait, Lebanon, Lithuania,  Malaysia, Mayotte, Mexico, Netherlands, Oman, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, San Marino, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Swaziland, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Vietnam",
      "recall_number": "Z-1423-2020",
      "product_description": "Triglycerides Reagent (TG 2 x 300), REF: 445850, For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233228",
      "product_quantity": "N/A",
      "reason_for_recall": "It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.",
      "recall_initiation_date": "20190815",
      "center_classification_date": "20200302",
      "termination_date": "20240228",
      "report_date": "20200311",
      "code_info": "All Lot Numbers"
    }
  ]
}