  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76212",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "address_2": "N/A",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including Puerto Rico",
      "recall_number": "Z-1421-2017",
      "product_description": "IMMULITE 2000 Progesterone; Catalog Number: L2KPW2   L2KPW6; Siemens Material Number: 10381181, 10381170    Product Usage:  For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers - for the quantitative measurement of progesterone in serum as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.",
      "product_quantity": "11963",
      "reason_for_recall": "DHEA-S causes falsely elevated progesterone results.",
      "recall_initiation_date": "20170104",
      "center_classification_date": "20170309",
      "termination_date": "20190718",
      "report_date": "20170315",
      "code_info": "SMN: 10381181: L2KPW2 kit lot 453  L2KPW2 kit lot 454  L2KPW2 kit lot 455  L2KPW2 kit lot 456  L2KPW2 kit lot 457  L2KPW2 kit lot 458  L2KPW2 kit lot 459  L2KPW2 kit lot 460  L2KPW2 kit lot 462  L2KPW2 kit lot 463  L2KPW2 kit lot 464  L2KPW2 kit lot 465  L2KPW2 kit lot 467  L2KPW2 kit lot 468  L2KPW2 kit lot 470  L2KPW2 kit lot 471  L2KPW2 kit lot 472  L2KPW2 kit lot 474;  SMN: 10381170: L2KPW6 kit lot 453  L2KPW6 kit lot 454  L2KPW6 kit lot 455  L2KPW6 kit lot 456  L2KPW6 kit lot 457  L2KPW6 kit lot 458  L2KPW6 kit lot 459  L2KPW6 kit lot 460  L2KPW6 kit lot 461  L2KPW6 kit lot 463  L2KPW6 kit lot 464  L2KPW6 kit lot 465  L2KPW6 kit lot 466  L2KPW6 kit lot 467  L2KPW6 kit lot 469  L2KPW6 kit lot 470  L2KPW6 kit lot 471  L2KPW6 kit lot 472  L2KPW6 kit lot 474"
    }
  ]
}