{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Carlsbad",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65165",
      "recalling_firm": "Alphatec Spine, Inc.",
      "address_1": "5818 El Camino Real",
      "address_2": "N/A",
      "postal_code": "92008-8816",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-1421-2013",
      "product_description": "Leucadia Autolok\" Polyaxial Pedicle Screw System.    Intended to help provide correction, immobilization and stabilization of spinal segments.",
      "product_quantity": "1,814",
      "reason_for_recall": "Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia\" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial  screw head and rod may fail to provide a rigid construct under loading. To date there have been  no reports of patient injury, adve",
      "recall_initiation_date": "20130506",
      "center_classification_date": "20130528",
      "termination_date": "20140925",
      "report_date": "20130605",
      "code_info": "83010  85001  85002  83005-30   83005-35   83005-40   83005-45   83005-50   83006-30   83006-35   83006-40   83006-45   83006-50   83006-55   83006-60   83007-30   83007-35   83007-40   83007-45   83007-50   83007-55   83008-30   83008-35   83008-40   83008-45   83008-50   83008-55   83008-80   82001-1       Catalog Numbers:  LA-HSS0010   LA-HDOO01   LA-CTOO02   LA-PS0530   LA-PS0535   LA-PS0540   LA-PS0545   LA-PS0550   LA-PS0630   LA-PS0635   LA-PS0640   LA-PS0645   LA-PS0650   LA-PS0655   LA-PS0660   LA-PS0730   LA-PS0735   LA-PS0740   LA-PS0745   LA-PS0750   LA-PS0755   LA-PS0830   LA-PS0835   LA-PS0840   LA-PS0845   LA-PS0850   LA-PS0855   LA-PS0880   LA-SOO01     Lots Affected: All"
    }
  ]
}