  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Columbia",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90488",
      "recalling_firm": "Ambu Inc.",
      "address_1": "6230 Old Dobbin Ln Ste 250",
      "address_2": "N/A",
      "postal_code": "21045-5955",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Domestic distribution nationwide.   Foreign distribution to Austria, Belgium, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Netherlands, Norge, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, UK  Canada, Australia, New Zealand, Israel  .",
      "recall_number": "Z-1420-2022",
      "product_description": "Ambu VivaSight 2 DLT; endobronchial tube system used for verifying tube placement and repositioning. Catalog numbers 412351000,   412371000,   412391000,   412411000",
      "product_quantity": "12,921",
      "reason_for_recall": "Firm has received complaints on the device concerning leak/rupture of either the bronchial or tracheal cuff. Leakage during procedure could lead to loss of secure airway and will require re-intubation of patient.",
      "recall_initiation_date": "20220603",
      "center_classification_date": "20220720",
      "report_date": "20220727",
      "code_info": "Catalog No: 412351000  UDI-DI: 5707480145706   1 each   5707480145713   5 pc/pack    Catalog No. 412371000  UDI-DI: 5707480145737   1 each  5707480145744   5 pc/pack    Catalog No. 412391000  UDI-DI: 5707480145768   1 each  5707480145775   5 pc/pack    Catalog No. 412411000  UDI-DI: 5707480145799   1 each  5707480145805   5 pc/pack      All lots"
    }
  ]
}