      "status": "Ongoing",
      "city": "Alfreton",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79627",
      "recalling_firm": "Autochair",
      "address_1": "Wood St North",
      "address_2": "Meadow Lane Industrial Estate",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "PA, and worldwide",
      "recall_number": "Z-1420-2018",
      "product_description": "Product: Milford Smartbase   Model #  9934105 Pride SmartBase  Pride product codes: LSMPLCG A & LSMPLCG B",
      "product_quantity": "",
      "reason_for_recall": "Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.",
      "recall_initiation_date": "20180307",
      "center_classification_date": "20180418",
      "report_date": "20180425",
      "code_info": "All Lots",
      "more_code_info": ""
    }
  ]
}