  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76212",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "address_2": "N/A",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including Puerto Rico",
      "recall_number": "Z-1420-2017",
      "product_description": "IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128    Product Usage:  For in vitro diagnostic use with the IMMULITE¿2000 Systems Analyzers- for the quantitative measurement of progesterone in serum, as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.",
      "product_quantity": "9076",
      "reason_for_recall": "DHEA-S causes falsely elevated progesterone results.",
      "recall_initiation_date": "20170104",
      "center_classification_date": "20170309",
      "termination_date": "20190718",
      "report_date": "20170315",
      "code_info": "LKPW1 kit lot 0245  LKPW1 kit lot 0246  LKPW1 kit lot 0247  LKPW1 kit lot 0248  LKPW1 kit lot 0249"
    }
  ]
}