  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91937",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "N/A",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of IL, NY and the countries of  Netherlands, Saudi Arabia.",
      "recall_number": "Z-1415-2023",
      "product_description": "BD Kiestra InoqulA+ with BD Kiestra\" InoqulA+\" and BeA Software version 5.1 and 5.1.1 as part of the Icefall A platform.  An in vitro diagnostic medical device intended to automate specimen processing according to user-defined procedures and protocols.  Catalog (Ref) No. : 447213",
      "product_quantity": "7 instruments",
      "reason_for_recall": "Upon installation of BD Kiestra\" InoqulA+\" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys\" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+\" solution is indicated for use in the clinical laboratory.",
      "recall_initiation_date": "20230322",
      "center_classification_date": "20230420",
      "report_date": "20230426",
      "code_info": "UDI: 038290LSUVLPKFPV Serial Numbers: INO-000156, INO-000249, INO-000248, INO-000252, INO-000254, 20130276, 19033ST002"
    }
  ]
}