  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87622",
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "address_1": "1080 US Highway 202 S",
      "address_2": "N/A",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  WorldWide",
      "recall_number": "Z-1414-2021",
      "product_description": "cobas Liat Analyzer: respiratory virus panel nucleic acid assay system  Catalog Number: 07341920190",
      "product_quantity": "US: 5,430 units ; OUS 1528 units",
      "reason_for_recall": "Invalid or false positive results for one or more targets (SARS-CoV-2/Influenza A/Influenza B) using the cobas SARS-CoV-2 & Influenza A/B assay for use under emergency use authorization on the cobas Liat analyzer, showed abnormal PCR growth curves",
      "recall_initiation_date": "20210325",
      "center_classification_date": "20210412",
      "report_date": "20210421",
      "code_info": "All serial ID numbers"
    }
  ]
}