  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sparks",
      "address_1": "7 Loveton Cir",
      "reason_for_recall": "The information in the EpiCenter interface translation tables was not updated to match the Bruker MALDI BioTyper database identifications for some organisms. This could lead to a possible mis-identification of an organism.",
      "address_2": "BD Diagnostic Systems",
      "product_quantity": "Qty Distr. to Field - 115",
      "code_info": "Lot Numbers  Version 6.20A    Cat. Number  442953",
      "center_classification_date": "20170308",
      "distribution_pattern": "15 US customers, 18 Ex-US sites, and 0 distributors.  Approximately 115 customers totally  and 0 distributors",
      "state": "MD",
      "product_description": "EpiCenter Plus Microbiology Data Management System software version 6.20A, when used in conjunction with the MALDI BioTyper System .    Product Usage:  The EpiCenter system is a clinical data management system designed for use with certain BD instrument systems. The EpiCenter system provides advanced capabilities in data management, including extended demographic data, comprehensive reporting features including a powerful database query tool, bi-directional Laboratory Information System (LIS) communications and long-term data storage. Additionally, the EpiCenter system provides an Expert System that can manipulate and/or act upon AST results.",
      "report_date": "20170315",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Co.",
      "recall_number": "Z-1414-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76498",
      "termination_date": "20170308",
      "recall_initiation_date": "20140825",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}