      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76456",
      "recalling_firm": "Encore Medical, Lp",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany",
      "recall_number": "Z-1413-2017",
      "product_description": "RSP Impaction Fixture",
      "product_quantity": "626 units",
      "reason_for_recall": "During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.",
      "recall_initiation_date": "20170209",
      "center_classification_date": "20170308",
      "termination_date": "20171020",
      "report_date": "20170315",
      "code_info": "109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02"
    }
  ]
}