{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76456",
      "recalling_firm": "Encore Medical, Lp",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany",
      "recall_number": "Z-1412-2017",
      "product_description": "Turon Impaction Fixture",
      "product_quantity": "297 units",
      "reason_for_recall": "During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.",
      "recall_initiation_date": "20170209",
      "center_classification_date": "20170308",
      "termination_date": "20171020",
      "report_date": "20170315",
      "code_info": "101521L01, 115670L18, 128092L10, 136172L07, 136172L08, 141511L22, 37369L10A, 37369L10B, 37837L23A, 37837L23B, 37837L23C, 37837L23D, 51103L05E, 51103L05F, 52748L02A, 52748L02B, 52748L02C, 52748L02D, 52748L02E, 52748L02F, 52748L02G, 52748L02H, 57074L05A, 89910L10"
    }
  ]
}