  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Seattle",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70810",
      "recalling_firm": "Natus Medical Incorporated",
      "address_1": "5900 1st Ave S",
      "address_2": "N/A",
      "postal_code": "98108-3248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "distributed worldwide",
      "recall_number": "Z-1412-2015",
      "product_description": "Natus neoBLUE blanket LED Phototherapy system which includes the following items:    006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad    006254 neoBLUE blanket LED Light Box    006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad    007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR    007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR      007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK    007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK    007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS    007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS",
      "product_quantity": "2611 units (1911 units in the US and 700 units outside the US)",
      "reason_for_recall": "neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.",
      "recall_initiation_date": "20150316",
      "center_classification_date": "20150410",
      "termination_date": "20161220",
      "report_date": "20150422",
      "code_info": "None"
    }
  ]
}