  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Leesburg",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87565",
      "recalling_firm": "K2M, Inc",
      "address_1": "600 Hope Pkwy SE",
      "address_2": "N/A",
      "postal_code": "20175-4428",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "PA",
      "recall_number": "Z-1410-2021",
      "product_description": "Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine)  Catalog number: 3608-90033",
      "product_quantity": "1 unit",
      "reason_for_recall": "Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies",
      "recall_initiation_date": "20210316",
      "center_classification_date": "20210412",
      "termination_date": "20210812",
      "report_date": "20210421",
      "code_info": "Lot Number: MNAJ  GTIN: 10888857082571"
    }
  ]
}