  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Aliso Viejo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84727",
      "recalling_firm": "Microvention, Inc.",
      "address_1": "35 Enterprise",
      "address_2": "N/A",
      "postal_code": "92656-2601",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "All distribution was OUS.  The countries receiving affected product are:  Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1407-2020",
      "product_description": "MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V,  100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V,  100254HS-V, 100256HS-V, 100304HS-V, 100306HS-V, 100308HS-V, 100406HS-V, 100408HS-V, and 100508HS-V.    Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.    The firm name on the label is MicroVention, Inc., Tustin, CA.",
      "product_quantity": "5,777 units",
      "reason_for_recall": "The devices may be missing the implant coil.",
      "recall_initiation_date": "20191122",
      "center_classification_date": "20200227",
      "termination_date": "20220126",
      "report_date": "20200304",
      "code_info": "Catalog/lot numbers:    100101HS-V - 1903045UH, 1903065UH, 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1905025UH, 1905085UH, 1905105ZH, 1905135UH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1903135WH, 1904175WH, 1905085WH, 1905105PH, and 1905135WH;    100102HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1903225ZH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905205ZH, 1905225UH, 1905225ZH, 1904035WH, and 1905155WH;    100103HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1904105UH, 1904155UH, 1904175UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, and 1905085WH;    100104HS-V - 1903115UH, 1903135UH, 1903225UH, 1904085UH, 1904105UH, 1904155UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1904175WH, 1904225WH, and 1905155WH;    100151HS-V - 1905025UH, 1903135TH, 1903085WH, 1903205WH, 1904015WH, 1904265WH, 1904295WH, 1905085WH, 1905155WH, and 1905225PH;    100152HS-V - 1903155UH, 1903185UH, 1903205UH, 1904085UH, 1904105UH, 1904155UH, 1904225UH, 1904245UH, 1905025UH, 1905025ZH, 1905035UH, 1905035ZH, 1905065UH. 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205ZH, 1903065NH, 1904175TH, 1904225NH, 1905065NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904085WH, 1904155WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155WH, and 1905225WH;    100153HS-V - 1903225UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1905225UH, 1905225ZH, 1903015NH, 1903065WH, 1903085WH, 1903205WH, 1904015WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225PH, 1904225WH, 1904245WH, 1904265WH, 1905025WH, 1905035PH, 1905065WH, 1905085WH, 1905155WH, 1905205PH, 1905225PH, and 1905225WH.    100154HS-V - 1903155UH, 1903185UH, 1903205UH, 1903225UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905155ZH, 1905175UH, 1905205UH, 1905225UH, 1903115TH, 1903015WH, 1903045WH, 1903065WH, 1904035WH, 1904055WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035PH, 1905035WH, 1905065WH, 1905085WH, 1905155PH, 1905155WH, 1905175WH, and 1905205WH;    100201HS-V - 1903135UH    100202HS-V - 1904085UH, 1904105UH, 1904175UH, 1904245UH, 1904225NH, 1905065NH, 1905205NH, 1903085WH, 1904015WH, 1904055WH, 1904175WH, 1904245WH, 1904265WH, 1904295WH, 1905085WH, 1905225PH, and 1905225WH;    100203HS-V - 1903155UH, 1903185UH, 1903225UH, 1904225NH, 1905135NH, 1903015WH, 1903065WH, 1904035WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905035WH, 1905155WH, and 1905225WH;    100204HS-V - 1904085UH, 1904105UH, 1904225UH, 1904245UH, 1905065UH, 1905085UH, 1905105UH, 1905135UH, 1905135ZH, 1905155UH, 1905175UH, 1905205UH, 1903045WH, 1903065WH, 1904015WH, 1904035WH, 1904155WH, 1904175WH, 1904225PH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035WH, 1905085WH, 1905155WH, and 1905225WH;    100206HS-V - 1904085UH, 1904105UH, 1904155UH, 1904175UH, 1904265UH, 1905065UH, 1905085UH, 1905105UH, 1905105ZH, 1905135UH, 1905155UH, 1905175UH, 1905205UH 1905205ZH, 1905225UH, 1903065WH, 1903085WH, 1903185WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, 1905065WH, 1905155WH, 1905175WH, and 1905225WH;     100254HS-V - 1903205WH, 1904015WH, 1904225WH, 1904265WH, 1904295WH, 1905025WH, 1905085WH, and 1905225PH;    100256HS-V - 1903065WH, 1903185WH, 1904015WH, and 1905155WH;    100304HS-V - 1903085NH, 1903135TH, 1904035TH, 1903065WH, 1905035PH, and 1905065WH;    100306HS-V - 1903115UH, 1903135UH, 1903155UH, 1903185UH, 1904085UH, 1903065WH, 1904035WH, 1904055WH, 1904155WH, 1904175WH, 1904225WH, 1904245WH, 1905035PH, 1905065WH, 1905205PH, and 1905225PH;    100308HS-V - 1903185UH, 1903225UH, 1905135UH, 1905175UH, 1905205UH, 1903085NH, 1903015WH, 1903065WH, 1903185WH, 1904015WH, 1904035WH, 1904055WH, 1904175WH, 1904225WH, 1904245WH, 1904265WH, 1904295WH, 1905025WH, 1905035PH, 1905035WH, and 1905065WH;    100406HS-V - 1903085WH, 1903185WH, 1904035WH, 1904225WH, and 1905205WH;    100408HS-V - 1905085UH, 1905105UH, 1905135UH, 1905155UH, 1903115TH, 1903015WH, 1903065WH, 1904265WH, 1904295WH, 1905085WH, 1905135WH, 1905155WH, and 1905205PH;    100508HS-V - 1903225TH, 1905205NH, 1903015WH, 1903085WH, 1903135WH, 1903185WH, 1904265WH, 1905105WH, and 1905205WH."
    }
  ]
}