  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.",
      "address_2": "",
      "product_quantity": "591",
      "code_info": "Serial no: 320309  336076  32104  336092  336150  336100  336228  336089  336095  336100  336116  32125  32123  336117  336120  336126  336139  336131  336149  336142  336150  336152  336161  336166  336150  336182  336203  336211  336217  336220  336221  336289  336283  336133  336145  336191  30035  345218  52073  336213  320227  320419  336188  32106  336123  336216  32117  32119  32151  66769829  320265  32144  32150  32160  32157  32159  32127  320363  32145  32129  32141  32133  32127  32168  336737  336760  336764  336766  336807  336809  320260  320268  320270  320271  320253  320276  320280  320285  320289  320291  320292  320318  320294  320316  320259  320279  320286  320326  320293  320327  320296  320290  320306  320321  320307  320311  320319  320315  320366  320390  320379  320371  320386  320323  320325  320340  320334  320329  320331  320391  320359  320384  320343  320365  320403  320374  320380  320393  320420  336800  32108  32096  32122  336802  32115  336803  32116  32114  32118  32121  32120  32126  32130  32131  32097  32129  32132  32133  32137  32134  32135  32136  32138  32142  32149  32154  32143  32145  32141  32140  336733  336734  32146  32147  32152  32161  32153  336735  32169  32148  336736  32156  32158  336738  32155  32160  336739  336740  336741  336742  32091  336743  336744  32165  336745  32163  336748  336746  32164  336747  32166  32167  32180  336749  336730  336776  336752  336777  32171  32173  336753  336754  32172  336757  336759  336756  336755  336758  336761  336762  336763  336765  336767  336768  336769  336772  336770  336756  336773  336775  336774  336779  336783  336785  336787  336786  336788  336789  336791  336793  336790  336795  336794  336799  336792  336801  336798  336796  336797  336805  336806  336808  336167  336266  32111  336104  336169  336261  32042  32111  336207  336179  336180  336267  336012  336115  3202694  320235  32179  32170  320201  320200  320220  320206  320221  320222  320262  320231  320237  320236  320254  320269  320245  320417  320275  320281  320303  320301  320288  320302  320370  320310  320313  320317  320344  320412  320367  320444  320381  320399  320411  320438  320429  320424  320358  320212  336096  320229  320266  336205  32103  336083  336115  336151  336125  32124  336147  336174  336196  336197  336212  336208  336242  336235  336253  336244  336257  336250  336265  336291  336270  336272  336288  336165  336101  336138  336151  336132  336189  336190  336215  336249  336277  320348  30025  32100  336068  32047  336128  336163  336144  32101  336246  32128  336129  336103  336067  336070  336071  336069  32101  336028  336144  336118  336112  336109  336130  336141  336286  336184  336183  336222  320239  336223  320239  336292  320335  336172  336206  336225  336227  336260  336263  336273  320099  32049  336136  336209  336153  336193  336195  336177  336285  320204  32174  336088  336106  32112  336134  336111  336119  336198  336122  336136  336137  336135  336140  336146  336155  336156  336159  336162  336181  336187  336251  336210  336214  336224  336230  336232  336243  336252  336262  336264  336287  336127  32181  336157  336218  336239  336256  336268  336282  32018  336093  336046  320342  336199  336086  336110  336108  336204  336091  336171  320238  320278  320304  320376  320432  320428  320352  336154  336160  336284  320351  320351  320215  336233  336173  336236  336170  320336  336102  32057  336219  336231  336226  32105  336085  32110  336158  320230  336168  320230  320410  336176  336087  336094  336098  336097  336124  336186  336238  336240  336255  336269  336113  32055  320237  320229  320236  336259  336237  336202  336194  336114  320355  32113  336234  336241  336258  336271  336254  336090  336105  336107  336164  336192  336175  336200  336201  336274  336247  336245  336248  336275  336276  336278  336290  336293  336279  336229  336178  32139  320228  320203  320202  320208  320219  320226  320223  320242  320284  320243  320282  320240  320257  320252  320272  320273  320274  320300  320345  320416  320377  320305  320207  320232  320216  320217  320209  320258  320356  320256  320261  320267  320312  320314  320320  320373  320360  320210  320213  320225  320218  320224  320233  320241  320234  320250  320263  320251  320277  320248  320255  320283  320295  320299  320362  320287  320357  320308  320350  320387  320392  320383  320332  336125  336185  336152  32147  32146  320229  32110  336751  336771  336815  336810  336812  336811  336813  336814  336819  336817  336816  336818  336820  336823  336821  336822  336824",
      "center_classification_date": "20180417",
      "distribution_pattern": "Worldwide Distribution - U.S. Nationwide. International distribution to Canada, Mexico, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cuba, Czech Republic, Denmark, Djibouti, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Laos, Lebanon, Lithuania, Macedonia, Malaysia, Maldives, Mali, Martinique, Monaco, Mongolia, Myanmar, Nepal, Netherlands, New Calcedonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saint Pierre and Miquelon, Samoa, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.",
      "state": "OH",
      "product_description": "Ingenuity Core 128 - Model no. 728323, 728324, Computed Tomography X-Ray system    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.",
      "report_date": "20180425",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-1407-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79666",
      "termination_date": "20200409",
      "more_code_info": "",
      "recall_initiation_date": "20180208",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}