  "results": [
    {
      "status": "Ongoing",
      "city": "Richardson",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92051",
      "recalling_firm": "GRAMERCY EXTREMITY ORTHOPEDICS",
      "address_1": "1239 N Glenville Dr",
      "address_2": "N/A",
      "postal_code": "75081-2412",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.",
      "recall_number": "Z-1395-2023",
      "product_description": "GEO Instrument Kit for EPS Plate, REF: 70120001, part of the GEO FirstFUSE 1st MTP Joint Arthrodesis Plating System",
      "product_quantity": "24",
      "reason_for_recall": "Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.",
      "recall_initiation_date": "20230404",
      "center_classification_date": "20230417",
      "report_date": "20230426",
      "code_info": "UDI-DI B471701200010, Serial Number(Expiration Date):  1022000493(10/13/27), 1022000623(10/13/27), 1022000683(10/13/27), 1022000243(10/13/27, 1022000313(10/13/27), 1022000913(10/13/27), 1022001213(10/13/27), 1022000403(10/13/27), 1022000383(10/13/27), 1022000883(10/13/27), 1022000353(10/13/27), 1022000663(10/13/27), 1022000733(10/13/27), 1022000793(10/13/27), 1022000983(10/13/27), 1022000993(10/13/27), 1022000373(10/13/27), 1022001193(10/13/27), 1022000523(10/13/27), 1022000283(10/13/27), 1022000223(10/13/27), 1022000363(10/13/27), 1022000413(10/13/27), 1022000453(10/13/27)"
    }
  ]
}