      "status": "Ongoing",
      "city": "Princeton",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90457",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "1100 Campus Rd",
      "address_2": "",
      "postal_code": "08540-6650",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US nationwide, Australia, New Zealand, Canada, Hong Kong, Pakistan, Georgia, Saudi Arabia, Israel, South Africa, United Arab Emirates, Austria, Belgium, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Portugal, Switzerland, Spain, Sweden, Slovakia, and United Kingdom.",
      "recall_number": "Z-1395-2022",
      "product_description": "CereLink ICP Monitor; Catalog No. 826820, 826820P.  Intracranial pressure monitor.",
      "product_quantity": "1,210 units",
      "reason_for_recall": "Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message \"sensor or extension cable failure!\" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.",
      "recall_initiation_date": "20220622",
      "center_classification_date": "20220722",
      "report_date": "20220803",
      "code_info": "UDI-DI- 10381780533778; All serial numbers.",
      "more_code_info": ""
    }
  ]
}