      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76360",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to Canada, Armenia, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Spain, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Kenya, Kuwait, Lithuania, Mauritius, Netherlands, Norway, Oman, Poland, Reunion, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom    VA/DOD: Yes",
      "recall_number": "Z-1390-2017",
      "product_description": "Flexor Radial Access Set, Individual product is packaged in a Tyvek-film sterilizable outer pouch.  The Flexor¿ Radial Hydrophilic Introducer Access Set is intended to introduce diagnostic and interventional devices In radial artery access procedures",
      "product_quantity": "41,975",
      "reason_for_recall": "Devices using a specific hemostatic valve design, referred to as a blue valve or polyisoprene valve, have an increase in reports of blood loss associated with the devices which could result in delay in procedure and blood loss from patients.",
      "recall_initiation_date": "20170201",
      "center_classification_date": "20170302",
      "termination_date": "20180928",
      "report_date": "20170308",
      "code_info": "Catalog Identifier: KCFN-4.0(5.0,6.0,7.0)- / -RA-(S)-HC    Codes consist of Lot #, RPN #, Manufacture Date and Expiration Date:  4427564X KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017  4427564XX KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017  4427564XXX KCFN-4.0-18-13-RA-HC 8/21/2014 8/21/2017  4427566X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017  4427568X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017  4610803X KCFN-4.0-18-13-RA-HC 2014/03 3/1/2017  4954615 KCFN-4.0-18-13-RA-HC 6/1/2014 6/1/2017  5087583 KCFN-4.0-18-13-RA-HC 8/1/2014 8/1/2017  5087583X KCFN-4.0-18-13-RA-HC 8/1/2014 8/1/2017  5102705 KCFN-4.0-18-13-RA-HC 8/11/2014 8/11/2017  5102705X KCFN-4.0-18-13-RA-HC 8/11/2014 8/11/2017  5260392 KCFN-4.0-18-13-RA-HC 10/6/2014 10/6/2017  5260392X KCFN-4.0-18-13-RA-HC 10/6/2014 10/6/2017  4427454X KCFN-4.0-18-23-RA-HC 8/21/2014 8/21/2017  4427454XX KCFN-4.0-18-23-RA-HC 8/21/2014 8/21/2017  4427456X KCFN-4.0-18-23-RA-HC 2014/03 3/1/2017  4427458X KCFN-4.0-18-23-RA-HC 2014/03 3/1/2017  4667209X KCFN-4.0-18-23-RA-HC 8/5/2014 8/5/2017  4844427 KCFN-4.0-18-23-RA-HC 2014/04 4/1/2017  4900689 KCFN-4.0-18-23-RA-HC 5/1/2014 5/1/2017  5579987 KCFN-4.0-18-23-RA-HC 2/25/2015 2/25/2018  4427654X KCFN-4.0-18-7-RA-HC 8/8/2014 8/8/2017  4427656X KCFN-4.0-18-7-RA-HC 2014/03 3/1/2017  4667251X KCFN-4.0-18-7-RA-HC 8/5/2014 8/5/2017  4767849 KCFN-4.0-18-7-RA-HC 2014/03 3/1/2017  5122477 KCFN-4.0-18-7-RA-HC 8/8/2014 8/8/2017  5167205 KCFN-4.0-18-7-RA-HC 8/29/2014 8/29/2017  5167205X KCFN-4.0-18-7-RA-HC 8/29/2014 8/29/2017  5537258 KCFN-4.0-18-7-RA-HC 2/10/2015 2/10/2018  5580037 KCFN-4.0-18-7-RA-HC 2/10/2015 2/10/2018  5628941 KCFN-4.0-18-7-RA-HC 3/16/2015 3/16/2018  5688179 KCFN-4.0-18-7-RA-HC 4/9/2015 4/9/2018  5797339 KCFN-4.0-18-7-RA-HC 5/1/2015 5/1/2018  5956253 KCFN-4.0-18-7-RA-HC 7/14/2015 7/14/2018  6101759 KCFN-4.0-18-7-RA-HC 9/4/2015 9/4/2018  6183817 KCFN-4.0-18-7-RA-HC 10/1/2015 10/1/2018  6273554 KCFN-4.0-18-7-RA-HC 11/8/2015 11/8/2018  6351813 KCFN-4.0-18-7-RA-HC 12/11/2015 12/11/2018  4427728X KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017  4427728XX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017  4427728XXX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017  4427730X KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017  4427730XX KCFN-5.0-18-13-RA-HC 8/21/2014 8/21/2017  4427732X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427734X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427736X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427738X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427740X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427744X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427746X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427748X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427750X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427752X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427754X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427756X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427758X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427760X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427762X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427764X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427766X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427768X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427770X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427772X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427774X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4427776X KCFN-5.0-18-13-RA-HC 2014/03 3/1/2017  4667336X KCFN-5.0-18-13-RA-HC 8/5/2014 8/5/2017  4846949 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017  4846951 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017  4846953 KCFN-5.0-18-13-RA-HC 2014/04 4/1/2017  4933698 KCFN-5.0-18-13-RA-HC 5/1/2014 5/1/2017  4933700 KCFN-5.0-18-13-RA-HC 6/1/2014 6/1/2017  4933702 KCFN-5.0-18-13-RA-HC 6/1/2014 6/1/2017  5034189 KCFN-5.0-18-13-RA-HC 7/14/2014 7/14/2017  5034191 KCFN-5.0-18-13-RA-HC 7/16/2014 7/16/2017  5034193 KCFN-5.0-18-13-RA-HC 7/14/2014 7/14/2017  5034195 KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017  5102745 KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017  5102745X KCFN-5.0-18-13-RA-HC 7/21/2014 7/21/2017  5112123 KCFN-5.0-18-13-RA-HC 8/8/2014 8/8/2017  5112125 KCFN-5.0-18-13-RA-HC 8/8/2014 8/8/2017  5112127 KCFN-5.0-18-13-RA-HC 8/15/2014 8/15/2017  5167398 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017  5167400 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017  5167402 KCFN-5.0-18-13-RA-HC 8/29/2014 8/29/2017  5167404 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017  5167406 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017  5224131 KCFN-5.0-18-13-RA-HC 9/5/2014 9/5/2017  5224133 KCFN-5.0-18-13-RA-HC 9/15/2014 9/15/2017  5224135 KCFN-5.0-18-13-RA-HC 9/18/2014 9/18/2017  5253913 KCFN-5.0-18-13-RA-HC 10/20/2014 10/20/2017  5253915 KCFN-5.0-18-13-RA-HC 10/14/2014 10/14/2017  5253917 KCFN-5.0-18-13-RA-HC 10/20/2014 10/20/2017  5253919 KCFN-5.0-18-13-RA-HC 10/31/2014 10/31/2017  5425165 KCFN-5.0-18-13-RA-HC 12/9/2014 12/9/2017  5425167 KCFN-5.0-18-13-RA-HC 12/16/2014 12/16/2017  5425169 KCFN-5.0-18-13-RA-HC 12/16/2014 12/16/2017  5493731 KCFN-5.0-18-13-RA-HC 1/14/2015 1/14/2018  5493733 KCFN-5.0-18-13-RA-HC 1/27/2015 1/27/2018  5493735 KCFN-5.0-18-13-RA-HC 1/14/2015 1/14/2018  5580062 KCFN-5.0-18-13-RA-HC 2/10/2015 2/10/2018  5589721 KCFN-5.0-18-13-RA-HC 3/18/2015 3/18/2018  5589723 KCFN-5.0-18-13-RA-HC 4/6/2015 4/6/2018  5589725 KCFN-5.0-18-13-RA-HC 4/9/2015 4/9/2018  5651232 KCFN-5.0-18-13-RA-HC 4/9/2015 4/9/2018  5651234 KCFN-5.0-18-13-RA-HC 4/16/2015 4/16/2018  5651236 KCFN-5.0-18-13-RA-HC 4/16/2015 4/16/2018  5651238 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018  5703845 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018  5703847 KCFN-5.0-18-13-RA-HC 5/1/2015 5/1/2018  5703849 KCFN-5.0-18-13-RA-HC 5/18/2015 5/18/2018  5893889 KCFN-5.0-18-13-RA-HC 7/9/2015 7/9/2018  5893891 KCFN-5.0-18-13-RA-HC 7/8/2015 7/8/2018  5971098 KCFN-5.0-18-13-RA-HC 7/29/2015 7/29/2018  5994633 KCFN-5.0-18-13-RA-HC 8/19/2015 8/19/2018  5994635 KCFN-5.0-18-13-RA-HC 8/4/2015 8/4/2018  6025703 KCFN-5.0-18-13-RA-HC 8/4/2015 8/4/2018  6025705 KCFN-5.0-18-13-RA-HC 8/2/2015 8/2/2018  6025707 KCFN-5.0-18-13-RA-HC 8/2/2015 8/2/2018  6117310 KCFN-5.0-18-13-RA-HC 9/10/2015 9/10/2018  6117312 KCFN-5.0-18-13-RA-HC 9/17/2015 9/17/2018  6176097 KCFN-5.0-18-13-RA-HC 9/27/2015 9/27/2018  6176099 KCFN-5.0-18-13-RA-HC 10/14/2015 10/14/2018  6192470 KCFN-5.0-18-13-RA-HC 10/2/2015 10/2/2018  6192472 KCFN-5.0-18-13-RA-HC 10/2/2015 10/2/2018  6284394 KCFN-5.0-18-13-RA-HC 12/9/2015 12/9/2018  6284396 KCFN-5.0-18-13-RA-HC 12/9/2015 12/9/2018  4427778X KCFN-5.0-18-13-RA-S-HC 2014/03 3/1/2017  4427780X KCFN-5.0-18-13-RA-S-HC 8/21/2014 8/21/2017  4427780XX KCFN-5.0-18-13-RA-S-HC 8/21/2014 8/21/2017  4667340X KCFN-5.0-18-13-RA-S-HC 8/5/2014 8/5/2017  4893616 KCFN-5.0-18-13-RA-S-HC 2014/04 4/1/2017  4897225 KCFN-5.0-18-13-RA-S-HC 2014/04 4/1/2017  4991650 KCFN-5.0-18-13-RA-S-HC 6/1/2014 6/1/2017  5102747 KCFN-5.0-18-13-RA-S-HC 8/1/2014 8/1/2017  5102747X KCFN-5.0-18-13-RA-S-HC 8/1/2014 8/1/2017  5580064 KCFN-5.0-18-13-RA-S-HC 2/10/2015 2/10/2018  4427680X KCFN-5.0-18-23-RA-HC 8/21/2014 8/21/2017  4427684X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427686X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427688X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427690X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427692X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427694X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427696X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427696X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427698X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427700X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4427702X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4605309X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4605311X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4691483X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4691485X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4691487X KCFN-5.0-18-23-RA-HC 2014/03 3/1/2017  4893601 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017  4893603 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017  4893605 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017  4900715 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017  4900717 KCFN-5.0-18-23-RA-HC 2014/04 4/1/2017  4900719 KCFN-5.0-18-23-RA-HC 2014/05 5/1/2017  4905653 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017  4905655 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017  4905657 KCFN-5.0-18-23-RA-HC 6/1/2014 6/1/2017  5094891 KCFN-5.0-18-23-RA-HC 8/8/2014 8/8/2017  5094891 KCFN-5.0-18-23-RA-HC 8/8/2014 8/8/2017  5094893 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017  5102736 KCFN-5.0-18-23-RA-HC 7/21/2014 7/21/2017  5102736X KCFN-5.0-18-23-RA-HC 7/21/2014 7/21/2017  5122485 KCFN-5.0-18-23-RA-HC 8/11/2014 8/11/2017  5122487 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017  5122489 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017  5143028 KCFN-5.0-18-23-RA-HC 8/15/2014 8/15/2017  5143030 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017  5171213 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017  5171215 KCFN-5.0-18-23-RA-HC 8/28/2014 8/28/2017  5171217 KCFN-5.0-18-23-RA-HC 9/12/2014 9/12/2017  5253880 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017  5253882 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017  5253884 KCFN-5.0-18-23-RA-HC 10/3/2014 10/3/2017  5263587 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017  5263589 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017  5298730 KCFN-5.0-18-23-RA-HC 10/20/2014 10/20/2017  5298732 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017  5298732 KCFN-5.0-18-23-RA-HC 10/14/2014 10/14/2017  5298734 KCFN-5.0-18-23-RA-HC 10/31/2014 10/31/2017  5364979 KCFN-5.0-18-23-RA-HC 12/15/2014 12/15/2017  5364981 KCFN-5.0-18-23-RA-HC 12/11/2014 12/11/2017  5364983 KCFN-5.0-18-23-RA-HC 12/9/2014 12/9/2017  5421713 KCFN-5.0-18-23-RA-HC 11/25/2014 11/25/2017  5421715 KCFN-5.0-18-23-RA-HC 12/11/2014 12/11/2017  5421717 KCFN-5.0-18-23-RA-HC 12/9/2014 12/9/2017  5520711 KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018  5520711X KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018  5520713 KCFN-5.0-18-23-RA-HC 1/27/2015 1/27/2018  5600196 KCFN-5.0-18-23-RA-HC 3/16/2015 3/16/2018  5600198 KCFN-5.0-18-23-RA-HC 3/16/2015 3/16/2018  5600200 KCFN-5.0-18-23-RA-HC 4/8/2015 4/8/2018  5675240 KCFN-5.0-18-23-RA-HC 4/22/2015 4/22/2018  5675242 KCFN-5.0-18-23-RA-HC 4/22/2015 4/22/2018  5675244 KCFN-5.0-18-23-RA-HC 5/18/2015 5/18/2018  5839392 KCFN-5.0-18-23-RA-HC 6/17/2015 6/17/2018  5839394 KCFN-5.0-18-23-RA-HC 6/26/2015 6/26/2018  5842468 KCFN-5.0-18-23-RA-HC 6/29/2015 6/29/2018  5842470 KCFN-5.0-18-23-RA-HC 6/24/2015 6/24/2018  6019212 KCFN-5.0-18-23-RA-HC 7/31/2015 7/31/2018  6019214 KCFN-5.0-18-23-RA-HC 7/31/2015 7/31/2018  6091378 KCFN-5.0-18-23-RA-HC 8/30/2015 8/30/2018  6192461 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018  6192463 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018  6212495 KCFN-5.0-18-23-RA-HC 10/15/2015 10/15/2018  6212497 KCFN-5.0-18-23-RA-HC 10/20/2015 10/20/2018  6237778 KCFN-5.0-18-23-RA-HC 11/8/2015 11/8/2018  6237780 KCFN-5.0-18-23-RA-HC 11/3/2015 11/3/2018  6263210 KCFN-5.0-18-23-RA-HC 11/8/2015 11/8/2018  6263212 KCFN-5.0-18-23-RA-HC 11/25/2015 11/25/2018  6280180 KCFN-5.0-18-23-RA-HC 11/29/2015 11/29/2018  6354966 KCFN-5.0-18-23-RA-HC 12/9/2015 12/9/2018  NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017  NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017  NS5230210 KCFN-5.0-18-23-RA-HC 10/6/2014 10/6/2017  4427788X KCFN-5.0-18-7-RA-HC 8/21/2014 8/21/2017  4427788XX KCFN-5.0-18-7-RA-HC 8/21/2014 8/21/2017  4667356X KCFN-5.0-18-7-RA-HC 8/5/2014 8/5/2017  4765057 KCFN-5.0-18-7-RA-HC 2014/02 2/1/2017  4838340 KCFN-5.0-18-7-RA-HC 2014/04 4/1/2017  4893627 KCFN-5.0-18-7-RA-HC 2014/04 4/1/2017  5167315 KCFN-5.0-18-7-RA-HC 10/6/2014 10/6/2017  5167315X KCFN-5.0-18-7-RA-HC 10/6/2014 10/6/2017  5357936 KCFN-5.0-18-7-RA-HC 12/9/2014 12/9/2017  5467702 KCFN-5.0-18-7-RA-HC 12/16/2014 12/16/2017  5467702X KCFN-5.0-18-7-RA-HC 12/16/2014 12/16/2017  5493765 KCFN-5.0-18-7-RA-HC 1/13/2015 1/13/2018  5580117 KCFN-5.0-18-7-RA-HC 2/10/2015 2/10/2018  5634995 KCFN-5.0-18-7-RA-HC 3/18/2015 3/18/2018  5666595 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018  5688231 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018  5727259 KCFN-5.0-18-7-RA-HC 5/11/2015 5/11/2018  5736885 KCFN-5.0-18-7-RA-HC 4/15/2015 4/15/2018  5739813 KCFN-5.0-18-7-RA-HC 5/1/2015 5/1/2018  5745146 KCFN-5.0-18-7-RA-HC 5/19/2015 5/19/2018  5748864 KCFN-5.0-18-7-RA-HC 5/12/2015 5/12/2018  5751980 KCFN-5.0-18-7-RA-HC 5/1/2015 5/1/2018  6307647 KCFN-5.0-18-7-RA-HC 12/11/2015 12/11/2018  4427570X KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017  4427570XX KCFN-6.0-18-13-RA-HC 8/20/2014 8/20/2017  4427572X KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017  4427572XX KCFN-6.0-18-13-RA-HC 8/21/2014 8/21/2017  4427578X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427580X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427582X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427584X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427586X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427588X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427590X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427594X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427596X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427598X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427600X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427602X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427604X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427606X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427608X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427610X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427612X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427614X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427616X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427618X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427620X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427622X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427624X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427626X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427628X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427630X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427632X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427634X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427636X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427638X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427640X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427642X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427644X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427646X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427648X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427650X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4427652X KCFN-6.0-18-13-RA-HC 2014/03 3/1/2017  4667237X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017  4667241X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017  4667247X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017  4667249X KCFN-6.0-18-13-RA-HC 8/5/2014 8/5/2017  4893552 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017  4893554 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017  4893556 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017  4893558 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017  4893562 KCFN-6.0-18-13-RA-HC 2014/04 4/1/2017  4893562X KCFN-6.0-18-13-RA-HC 4/1/2014 4/1/2017  5005398 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017  5005844 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017  5005846 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017  5005848 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017  5005850 KCFN-6.0-18-13-RA-HC 6/1/2014 6/1/2017  5005852 KCFN-6.0-18-13-RA-HC 7/1/2014 7/1/2017  5034172 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017  5034174 KCFN-6.0-18-13-RA-HC 7/14/2014 7/14/2017  5034176 KCFN-6.0-18-13-RA-HC 7/15/2014 7/15/2017  5034178 KCFN-6.0-18-13-RA-HC 7/14/2014 7/14/2017  5070910 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017  5070912 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017  5070914 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5070916X KCFN-6.0-18-13-RA-HC 8/8/2014 8/8/2017  5070918 KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017  5102709 KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017  5102709X KCFN-6.0-18-13-RA-HC 8/1/2014 8/1/2017  5102711 KCFN-6.0-18-13-RA-HC 7/24/2014 7/24/2017  5102713 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5102713X KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5102715 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5102715X KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5112129 KCFN-6.0-18-13-RA-HC 8/11/2014 8/11/2017  5112131 KCFN-6.0-18-13-RA-HC 8/15/2014 8/15/2017  5112133 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017  5112135 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017  5152643 KCFN-6.0-18-13-RA-HC 9/18/2014 9/18/2017  5152645 KCFN-6.0-18-13-RA-HC 9/18/2014 9/18/2017  5152647 KCFN-6.0-18-13-RA-HC 9/15/2014 9/15/2017  5152649 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5152651 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5152653 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5197821 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5197823 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5197825 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017  5197827 KCFN-6.0-18-13-RA-HC 10/3/2014 10/3/2017  5197829 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017  5219648 KCFN-6.0-18-13-RA-HC 10/14/2014 10/14/2017  5219650 KCFN-6.0-18-13-RA-HC 10/31/2014 10/31/2017  5219652 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017  5219654 KCFN-6.0-18-13-RA-HC 10/31/2014 10/31/2017  5219656 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017  5219658 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017  5289239 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017  5289241 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017  5289243 KCFN-6.0-18-13-RA-HC 11/10/2014 11/10/2017  5289245 KCFN-6.0-18-13-RA-HC 11/6/2014 11/6/2017  5289247 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017  5289247 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017  5322547 KCFN-6.0-18-13-RA-HC 11/10/2014 11/10/2017  5322549 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5322551 KCFN-6.0-18-13-RA-HC 11/11/2014 11/11/2017  5322553 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5350870 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5350872 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5350874 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5350876 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5350878 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5373504 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5373506 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017  5373508 KCFN-6.0-18-13-RA-HC 11/25/2014 11/25/2017  5373510 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017  5373512 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017  5432906 KCFN-6.0-18-13-RA-HC 12/9/2014 12/9/2017  5432908 KCFN-6.0-18-13-RA-HC 12/16/2014 12/16/2017  5432910 KCFN-6.0-18-13-RA-HC 1/8/2015 1/8/2018  5432912 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018  5449003 KCFN-6.0-18-13-RA-HC 1/12/2015 1/12/2018  5449005 KCFN-6.0-18-13-RA-HC 1/12/2015 1/12/2018  5449007 KCFN-6.0-18-13-RA-HC 1/6/2015 1/6/2018  5449009 KCFN-6.0-18-13-RA-HC 12/31/2014 12/31/2017  5449011 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018  5478604 KCFN-6.0-18-13-RA-HC 1/7/2015 1/7/2018  5478606 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5478608 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5478610 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5478612 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5511396 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5511396X KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5511398 KCFN-6.0-18-13-RA-HC 1/14/2015 1/14/2018  5511400 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5511402 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5511404 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5511406 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5533811 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5543747 KCFN-6.0-18-13-RA-HC 1/27/2015 1/27/2018  5543751 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018  5543753 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018  5543755 KCFN-6.0-18-13-RA-HC 2/25/2015 2/25/2018  5569070 KCFN-6.0-18-13-RA-HC 2/25/2015 2/25/2018  5569072 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018  5569074 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018  5569074X KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018  5569076 KCFN-6.0-18-13-RA-HC 3/19/2015 3/19/2018  5569078 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018  5569080 KCFN-6.0-18-13-RA-HC 4/6/2015 4/6/2018  5580025 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018  5580027 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018  5580029 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018  5580031 KCFN-6.0-18-13-RA-HC 2/10/2015 2/10/2018  5580033 KCFN-6.0-18-13-RA-HC 2/12/2015 2/12/2018  5597115 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018  5597117 KCFN-6.0-18-13-RA-HC 3/18/2015 3/18/2018  5597119 KCFN-6.0-18-13-RA-HC 4/9/2015 4/9/2018  5597121 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018  5597121X KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018  5597123 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018  5615321 KCFN-6.0-18-13-RA-HC 3/30/2015 3/30/2018  5632276 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5648402 KCFN-6.0-18-13-RA-HC 4/8/2015 4/8/2018  5648404 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5648406 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018  5648408 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018  5648410 KCFN-6.0-18-13-RA-HC 4/20/2015 4/20/2018  5648412 KCFN-6.0-18-13-RA-HC 4/16/2015 4/16/2018  5666544 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5666546 KCFN-6.0-18-13-RA-HC 4/21/2015 4/21/2018  5666548 KCFN-6.0-18-13-RA-HC 4/21/2015 4/21/2018  5666550 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5666552 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5684510 KCFN-6.0-18-13-RA-HC 2/28/2015 2/28/2018  5717938 KCFN-6.0-18-13-RA-HC 5/1/2015 5/1/2018  5717940 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018  5717942 KCFN-6.0-18-13-RA-HC 5/12/2015 5/12/2018  5717944 KCFN-6.0-18-13-RA-HC 5/12/2015 5/12/2018  5717946 KCFN-6.0-18-13-RA-HC 5/19/2015 5/19/2018  5758608 KCFN-6.0-18-13-RA-HC 5/26/2015 5/26/2018  5758610 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018  5758612 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018  5758614 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018  5758616 KCFN-6.0-18-13-RA-HC 5/22/2015 5/22/2018  5797329 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5797331 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018  5797333 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5797335 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5797337 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5825850 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018  5825852 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018  5825854 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5825856 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5825858 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5845959 KCFN-6.0-18-13-RA-HC 6/19/2015 6/19/2018  5845961 KCFN-6.0-18-13-RA-HC 6/17/2015 6/17/2018  5845963 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018  5845965 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018  5893856 KCFN-6.0-18-13-RA-HC 6/24/2015 6/24/2018  5893858 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018  5893860 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018  5893862 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018  5893864 KCFN-6.0-18-13-RA-HC 7/8/2015 7/8/2018  5909225 KCFN-6.0-18-13-RA-HC 7/13/2015 7/13/2018  5909227 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5909229 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5909231 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5909233 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5909235 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5952767 KCFN-6.0-18-13-RA-HC 7/16/2015 7/16/2018  5952769 KCFN-6.0-18-13-RA-HC 7/15/2015 7/15/2018  5952771 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5952773 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5959457 KCFN-6.0-18-13-RA-HC 7/15/2015 7/15/2018  5959459 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5959461 KCFN-6.0-18-13-RA-HC 7/14/2015 7/14/2018  5959463 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5959465 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5984820 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5984822 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5984824 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5984826 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  5984828 KCFN-6.0-18-13-RA-HC 7/21/2015 7/21/2018  6032995 KCFN-6.0-18-13-RA-HC 8/18/2015 8/18/2018  6032997 KCFN-6.0-18-13-RA-HC 8/2/2015 8/2/2018  6032999 KCFN-6.0-18-13-RA-HC 8/19/2015 8/19/2018  6032999 KCFN-6.0-18-13-RA-HC 8/19/2015 8/19/2018  6033001 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018  6033003 KCFN-6.0-18-13-RA-HC 8/30/2015 8/30/2018  6109716 KCFN-6.0-18-13-RA-HC 9/4/2015 9/4/2018  6109718 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018  6109720 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018  6109722 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018  6109724 KCFN-6.0-18-13-RA-HC 9/10/2015 9/10/2018  6117265 KCFN-6.0-18-13-RA-HC 9/10/2015 9/10/2018  6117267 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018  6117269 KCFN-6.0-18-13-RA-HC 9/17/2015 9/17/2018  6117271 KCFN-6.0-18-13-RA-HC 9/16/2015 9/16/2018  6176082 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018  6176084 KCFN-6.0-18-13-RA-HC 10/1/2015 10/1/2018  6176086 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018  6176088 KCFN-6.0-18-13-RA-HC 10/2/2015 10/2/2018  6224299 KCFN-6.0-18-13-RA-HC 10/16/2015 10/16/2018  6224301 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018  6224303 KCFN-6.0-18-13-RA-HC 11/8/2015 11/8/2018  6224305 KCFN-6.0-18-13-RA-HC 11/3/2015 11/3/2018  6240884 KCFN-6.0-18-13-RA-HC 11/3/2015 11/3/2018  6240886 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018  6240888 KCFN-6.0-18-13-RA-HC 10/29/2015 10/29/2018  6240890 KCFN-6.0-18-13-RA-HC 11/8/2015 11/8/2018  6251597 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018  6251599 KCFN-6.0-18-13-RA-HC 11/29/2015 11/29/2018  6251601 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018  6251601X KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018  6251603 KCFN-6.0-18-13-RA-HC 11/25/2015 11/25/2018  6293425 KCFN-6.0-18-13-RA-HC 12/9/2015 12/9/2018  6293427 KCFN-6.0-18-13-RA-HC 12/9/2015 12/9/2018  4427704X KCFN-6.0-18-13-RA-S-HC 8/21/2014 8/21/2017  4427704XX KCFN-6.0-18-13-RA-S-HC 8/21/2014 8/21/2017  4427706X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427708X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427710X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427712X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427714X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427716X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427718X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427720X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427722X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4427724X KCFN-6.0-18-13-RA-S-HC 2014/03 3/1/2017  4667286X KCFN-6.0-18-13-RA-S-HC 8/5/2014 8/5/2017  4667292X KCFN-6.0-18-13-RA-S-HC 8/5/2014 8/5/2017  4897210 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017  4897212 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017  4900721 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017  4900723 KCFN-6.0-18-13-RA-S-HC 4/1/2014 4/1/2017  4900725 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017  4900727 KCFN-6.0-18-13-RA-S-HC 2014/04 4/1/2017  4905659 KCFN-6.0-18-13-RA-S-HC 7/1/2014 7/1/2017  4905661 KCFN-6.0-18-13-RA-S-HC 7/1/2014 7/1/2017  5005399 KCFN-6.0-18-13-RA-S-HC 6/1/2014 6/1/2017  5102738 KCFN-6.0-18-13-RA-S-HC 8/1/2014 8/1/2017  5102740 KCFN-6.0-18-13-RA-S-HC 8/11/2014 8/11/2017  5102740X KCFN-6.0-18-13-RA-S-HC 8/11/2014 8/11/2017  5520550 KCFN-6.0-18-13-RA-S-HC 1/14/2015 1/14/2018  5520552 KCFN-6.0-18-13-RA-S-HC 1/14/2015 1/14/2018  5520554 KCFN-6.0-18-13-RA-S-HC 1/27/2015 1/27/2018  5580057 KCFN-6.0-18-13-RA-S-HC 2/25/2015 2/25/2018  4427538X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4427540X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4427542X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4427546X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4427548X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556115X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556117X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556119X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556121X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556123X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4556123X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600056XX KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600058X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600060X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600062X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600064X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4600066X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4635509X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4635511X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4635513X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4635515X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4635517X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4676509X KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4718572 KCFN-6.0-18-23-RA-HC 2014/01 1/1/2017  4718572X KCFN-6.0-18-23-RA-HC 2014/01 1/1/2017  4718574 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017  4718576 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4718578 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4718580 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4770554 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017  4770556 KCFN-6.0-18-23-RA-HC 2014/02 2/1/2017  4770558 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4770560 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4770562 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4770564 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4770566 KCFN-6.0-18-23-RA-HC 2014/03 3/1/2017  4844431 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844431X KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844433 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844435 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844437 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844439 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844441 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4844443 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879430 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879432 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879434 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879436 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879438 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4879440 KCFN-6.0-18-23-RA-HC 2014/04 4/1/2017  4917588 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917590 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917592 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917594 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917596 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917598 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917600 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4917602 KCFN-6.0-18-23-RA-HC 5/1/2014 5/1/2017  4977711 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4977713 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4977715 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4977717 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4977719 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981344 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981346 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981348 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981350 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981352 KCFN-6.0-18-23-RA-HC 7/1/2014 7/1/2017  4981354 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  4981356 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008370 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008372 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008374 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008376 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008378 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008380 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008382 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008384 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5008386 KCFN-6.0-18-23-RA-HC 6/1/2014 6/1/2017  5042551 KCFN-6.0-18-23-RA-HC 7/14/2014 7/14/2017  5042553 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017  5042555 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017  5042557 KCFN-6.0-18-23-RA-HC 7/14/2014 7/14/2017  5042559 KCFN-6.0-18-23-RA-HC 7/15/2014 7/15/2017  5042561 KCFN-6.0-18-23-RA-HC 7/21/2014 7/21/2017  5070896 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5070898 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5070900 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5070902 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5070904 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5070906 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5076722 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5076724 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5076726 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5076728 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5076730 KCFN-6.0-18-23-RA-HC 8/1/2014 8/1/2017  5102703 KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5102703X KCFN-6.0-18-23-RA-HC 7/25/2014 7/25/2017  5112137 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017  5112139 KCFN-6.0-18-23-RA-HC 8/8/2014 8/8/2017  5112141 KCFN-6.0-18-23-RA-HC 8/11/2014 8/11/2017  5112143 KCFN-6.0-18-23-RA-HC 8/8/2014 8/8/2017  5112145 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5112147 KCFN-6.0-18-23-RA-HC 9/8/2014 9/8/2017  5126714 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017  5126716 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017  5126718 KCFN-6.0-18-23-RA-HC 8/11/2014 8/11/2017  5126720 KCFN-6.0-18-23-RA-HC 8/28/2014 8/28/2017  5126722 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017  5126724 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5126726 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5126728 KCFN-6.0-18-23-RA-HC 9/9/2014 9/9/2017  5152608 KCFN-6.0-18-23-RA-HC 8/15/2014 8/15/2017  5152610 KCFN-6.0-18-23-RA-HC 9/18/2014 9/18/2017  5152612 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017  5152614 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017  5152616 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017  5152618 KCFN-6.0-18-23-RA-HC 9/8/2014 9/8/2017  5152620 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017  5152622 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5152624 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5152626 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5178902 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5178904 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5178906 KCFN-6.0-18-23-RA-HC 9/15/2014 9/15/2017  5178908 KCFN-6.0-18-23-RA-HC 9/12/2014 9/12/2017  5178910 KCFN-6.0-18-23-RA-HC 10/3/2014 10/3/2017  5178912 KCFN-6.0-18-23-RA-HC 10/3/2014 10/3/2017  5210537 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017  5210539 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017  5210541 KCFN-6.0-18-23-RA-HC",
      "more_code_info": "10/31/2014 10/31/2017  5210543 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017  5210545 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5210547 KCFN-6.0-18-23-RA-HC 10/31/2014 10/31/2017  5210549 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5244230 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5244232 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5244234 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5244236 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5244238 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017  5244240 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5263531 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5263531X KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5263533 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5263535 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017  5263535 KCFN-6.0-18-23-RA-HC 11/10/2014 11/10/2017  5263537 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5263539 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5263541 KCFN-6.0-18-23-RA-HC 11/11/2014 11/11/2017  5325914 KCFN-6.0-18-23-RA-HC 11/6/2014 11/6/2017  5325916 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017  5325918 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017  5325920 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5325922 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5347661 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5347663 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5347665 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5347667 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017  5347669 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5347671 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5376572 KCFN-6.0-18-23-RA-HC 11/25/2014 11/25/2017  5376574 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5376576 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5376578 KCFN-6.0-18-23-RA-HC 11/24/2014 11/24/2017  5376580 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017  5376582 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017  5376584 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017  5394985 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017  5394987 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017  5394989 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017  5394991 KCFN-6.0-18-23-RA-HC 12/11/2014 12/11/2017  5394993 KCFN-6.0-18-23-RA-HC 12/9/2014 12/9/2017  5394995 KCFN-6.0-18-23-RA-HC 12/8/2014 12/8/2017  5394997 KCFN-6.0-18-23-RA-HC 12/11/2014 12/11/2017  5491309 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018  5491311 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018  5491313 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018  5491315 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018  5491317 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018  5491319 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018  5514280 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018  5514282 KCFN-6.0-18-23-RA-HC 1/27/2015 1/27/2018  5514284 KCFN-6.0-18-23-RA-HC 1/14/2015 1/14/2018  5514286 KCFN-6.0-18-23-RA-HC 1/12/2015 1/12/2018  5514288 KCFN-6.0-18-23-RA-HC 1/27/2015 1/27/2018  5514290 KCFN-6.0-18-23-RA-HC 1/23/2015 1/23/2018  5514292 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5547576 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5547578 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5547580 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5547582 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5547584 KCFN-6.0-18-23-RA-HC 2/10/2015 2/10/2018  5573174 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018  5573176 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018  5573178 KCFN-6.0-18-23-RA-HC 2/25/2015 2/25/2018  5573180 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018  5573182 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018  5573182X KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018  5573184 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018  5619200 KCFN-6.0-18-23-RA-HC 3/16/2015 3/16/2018  5619202 KCFN-6.0-18-23-RA-HC 3/30/2015 3/30/2018  5619204 KCFN-6.0-18-23-RA-HC 4/8/2015 4/8/2018  5619206 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5619206 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5619208 KCFN-6.0-18-23-RA-HC 3/30/2015 3/30/2018  5637359 KCFN-6.0-18-23-RA-HC 4/9/2015 4/9/2018  5637361 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5637363 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5637365 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5637367 KCFN-6.0-18-23-RA-HC 4/16/2015 4/16/2018  5637369 KCFN-6.0-18-23-RA-HC 4/22/2015 4/22/2018  5637371 KCFN-6.0-18-23-RA-HC 4/15/2015 4/15/2018  5703791 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018  5703793 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018  5703795 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018  5703797 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018  5703799 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018  5703801 KCFN-6.0-18-23-RA-HC 5/18/2015 5/18/2018  5703803 KCFN-6.0-18-23-RA-HC 5/19/2015 5/19/2018  5834809 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018  5834811 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018  5834813 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018  5834815 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018  5861895 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018  5861897 KCFN-6.0-18-23-RA-HC 6/26/2015 6/26/2018  5861899 KCFN-6.0-18-23-RA-HC 6/24/2015 6/24/2018  5861901 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018  5861903 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018  5861905 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018  5899836 KCFN-6.0-18-23-RA-HC 7/9/2015 7/9/2018  5899838 KCFN-6.0-18-23-RA-HC 7/9/2015 7/9/2018  5899840 KCFN-6.0-18-23-RA-HC 7/8/2015 7/8/2018  5899842 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018  5899844 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018  5916332 KCFN-6.0-18-23-RA-HC 7/14/2015 7/14/2018  5916334 KCFN-6.0-18-23-RA-HC 7/14/2015 7/14/2018  5916336 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018  5916338 KCFN-6.0-18-23-RA-HC 7/13/2015 7/13/2018  5988945 KCFN-6.0-18-23-RA-HC 7/30/2015 7/30/2018  5988947 KCFN-6.0-18-23-RA-HC 7/29/2015 7/29/2018  5988949 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018  5988951 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  5988953 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018  6004265 KCFN-6.0-18-23-RA-HC 7/29/2015 7/29/2018  6004267 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6004269 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018  6004271 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6004273 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6004275 KCFN-6.0-18-23-RA-HC 8/2/2015 8/2/2018  6011946 KCFN-6.0-18-23-RA-HC 7/31/2015 7/31/2018  6011948 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6011950 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6011952 KCFN-6.0-18-23-RA-HC 8/2/2015 8/2/2018  6011954 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6011956 KCFN-6.0-18-23-RA-HC 8/4/2015 8/4/2018  6011958 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018  6055589 KCFN-6.0-18-23-RA-HC 8/24/2015 8/24/2018  6055591 KCFN-6.0-18-23-RA-HC 8/17/2015 8/17/2018  6055593 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018  6055595 KCFN-6.0-18-23-RA-HC 8/18/2015 8/18/2018  6055597 KCFN-6.0-18-23-RA-HC 8/17/2015 8/17/2018  6055599 KCFN-6.0-18-23-RA-HC 8/30/2015 8/30/2018  6055601 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018  6080971 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018  6080973 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018  6080975 KCFN-6.0-18-23-RA-HC 9/4/2015 9/4/2018  6080977 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018  6080979 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018  6150471 KCFN-6.0-18-23-RA-HC 9/16/2015 9/16/2018  6150473 KCFN-6.0-18-23-RA-HC 9/25/2015 9/25/2018  6150475 KCFN-6.0-18-23-RA-HC 9/9/2015 9/9/2018  6150477 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018  6154447 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018  6154449 KCFN-6.0-18-23-RA-HC 9/18/2015 9/18/2018  6154451 KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018  6154451X KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018  6154453 KCFN-6.0-18-23-RA-HC 10/2/2015 10/2/2018  6183805 KCFN-6.0-18-23-RA-HC 9/27/2015 9/27/2018  6183807 KCFN-6.0-18-23-RA-HC 10/2/2015 10/2/2018  6183809 KCFN-6.0-18-23-RA-HC 10/4/2015 10/4/2018  6183811 KCFN-6.0-18-23-RA-HC 10/1/2015 10/1/2018  6217250 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217252 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217254 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217256 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217258 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217260 KCFN-6.0-18-23-RA-HC 10/15/2015 10/15/2018  6217262 KCFN-6.0-18-23-RA-HC 11/3/2015 11/3/2018  6266738 KCFN-6.0-18-23-RA-HC 11/8/2015 11/8/2018  6266740 KCFN-6.0-18-23-RA-HC 11/29/2015 11/29/2018  6266742 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018  6266744 KCFN-6.0-18-23-RA-HC 11/29/2015 11/29/2018  6266746 KCFN-6.0-18-23-RA-HC 12/11/2015 12/11/2018  6266746 KCFN-6.0-18-23-RA-HC 12/11/2015 12/11/2018  6266748 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018  6266750 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018  6280152 KCFN-6.0-18-23-RA-HC 11/25/2015 11/25/2018  6280154 KCFN-6.0-18-23-RA-HC 12/9/2015 12/9/2018  6280156 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018  6280156 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018  6280158 KCFN-6.0-18-23-RA-HC 12/8/2015 12/8/2018  6280160 KCFN-6.0-18-23-RA-HC 12/9/2015 12/9/2018  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5147585 KCFN-6.0-18-23-RA-HC 9/30/2014 9/30/2017  NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017  NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017  NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017  NS5230212 KCFN-6.0-18-23-RA-HC 11/13/2014 11/13/2017  NS6061690 KCFN-6.0-18-23-RA-HC 8/28/2015 8/28/2018  4427660X KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017  4427662X KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017  4427662XX KCFN-6.0-18-23-RA-S-HC 8/21/2014 8/21/2017  4427664X KCFN-6.0-18-23-RA-S-HC 2014/03 3/1/2017  4667267X KCFN-6.0-18-23-RA-S-HC 8/5/2014 8/5/2017  4793616 KCFN-6.0-18-23-RA-S-HC 2014/04 4/1/2017  4893580 KCFN-6.0-18-23-RA-S-HC 2014/04 4/1/2017  4905625 KCFN-6.0-18-23-RA-S-HC 6/1/2014 6/1/2017  6025691 KCFN-6.0-18-23-RA-S-HC 8/4/2015 8/4/2018  6025693 KCFN-6.0-18-23-RA-S-HC 7/31/2015 7/31/2018  6041168 KCFN-6.0-18-23-RA-S-HC 8/18/2015 8/18/2018  4427658X KCFN-6.0-18-7-RA-HC 8/21/2014 8/21/2017  4427658XX KCFN-6.0-18-7-RA-HC 8/21/2014 8/21/2017  4893578 KCFN-6.0-18-7-RA-HC 2014/04 4/1/2017  5102732 KCFN-6.0-18-7-RA-HC 7/25/2014 7/25/2017  5102732X KCFN-6.0-18-7-RA-HC 7/25/2014 7/25/2017  5147478 KCFN-6.0-18-7-RA-HC 9/15/2014 9/15/2017  5585602 KCFN-6.0-18-7-RA-HC 2/10/2015 2/10/2018  5700318 KCFN-6.0-18-7-RA-HC 5/1/2015 5/1/2018  5842464 KCFN-6.0-18-7-RA-HC 6/19/2015 6/19/2018  NS6035989 KCFN-6.0-18-7-RA-HC 8/18/2015 8/18/2018  4427452X KCFN-7.0-18-13-RA-HC 8/21/2014 8/21/2017  4427452XX KCFN-7.0-18-13-RA-HC 8/21/2014 8/21/2017  4667207X KCFN-7.0-18-13-RA-HC 8/5/2014 8/5/2017  4893510 KCFN-7.0-18-13-RA-HC 2014/05 5/1/2017  4893512 KCFN-7.0-18-13-RA-HC 2014/04 4/1/2017  5143024 KCFN-7.0-18-13-RA-HC 8/15/2014 8/15/2017  5529606 KCFN-7.0-18-13-RA-HC 1/27/2015 1/27/2018  5566469 KCFN-7.0-18-13-RA-HC 2/25/2015 2/25/2018  5579985 KCFN-7.0-18-13-RA-HC 2/10/2015 2/10/2018  5634985 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018  5644741 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018  5675202 KCFN-7.0-18-13-RA-HC 5/1/2015 5/1/2018  5855292 KCFN-7.0-18-13-RA-HC 7/14/2015 7/14/2018  6205756 KCFN-7.0-18-13-RA-HC 10/2/2015 10/2/2018  4427782X KCFN-7.0-18-23-RA-HC 8/21/2014 8/21/2017  4427784X KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017  4427786X KCFN-7.0-18-23-RA-HC 8/6/2014 8/6/2017  4556159X KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017  4667354X KCFN-7.0-18-23-RA-HC 8/5/2014 8/5/2017  4814134 KCFN-7.0-18-23-RA-HC 2014/03 3/1/2017  4905684 KCFN-7.0-18-23-RA-HC 5/1/2014 5/1/2017  5126734 KCFN-7.0-18-23-RA-HC 8/1/2014 8/1/2017  5147561 KCFN-7.0-18-23-RA-HC 8/28/2014 8/28/2017  5156356 KCFN-7.0-18-23-RA-HC 8/28/2014 8/28/2017  5306553 KCFN-7.0-18-23-RA-HC 10/14/2014 10/14/2017  5316191 KCFN-7.0-18-23-RA-HC 11/10/2014 11/10/2017  5354346 KCFN-7.0-18-23-RA-HC 11/6/2014 11/6/2017  5493761 KCFN-7.0-18-23-RA-HC 1/13/2015 1/13/2018  5562614 KCFN-7.0-18-23-RA-HC 2/10/2015 2/10/2018  5585643 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018  5641608 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018  5641610 KCFN-7.0-18-23-RA-HC 3/16/2015 3/16/2018  5644799 KCFN-7.0-18-23-RA-HC 5/18/2015 5/18/2018  5682869 KCFN-7.0-18-23-RA-HC 5/18/2015 5/18/2018  5684511 KCFN-7.0-18-23-RA-HC 2/28/2015 2/28/2018  5842502 KCFN-7.0-18-23-RA-HC 7/14/2015 7/14/2018  6012026 KCFN-7.0-18-23-RA-HC 8/4/2015 8/4/2018  6188966 KCFN-7.0-18-23-RA-HC 10/1/2015 10/1/2018  6284401 KCFN-7.0-18-23-RA-HC 11/29/2015 11/29/2018"
    }
  ]
}