  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Addison",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73054",
      "recalling_firm": "Arjo, Inc. dba ArjoHuntleigh",
      "address_1": "2349 W Lake St",
      "address_2": "N/A",
      "postal_code": "60101-6183",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Consignees; 222 units distributed in the following states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and WV.  ***Foreign Consignees; 1374 units distributed in the following countries:  Australia, Austria, Belgium, Bosnia & Hercegovina, Brazil, CANADA, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, Iceland, Ireland, Italy, Japan, Lithuania, Luxemburg, Netherlands, New Zealand, Nigeria, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan and  UK.",
      "recall_number": "Z-1388-2016",
      "product_description": "The Concerto Basic Shower Trolley is intended for assisted hygiene care, especially showering and bathing of residents in care environments such as senior/assisted living, group home, special care, nursing homes, hospitals and home care. The Concerto Basic Shower Trolley is intended for indoor use.",
      "product_quantity": "1,596 units",
      "reason_for_recall": "Reports of the bolts connecting the stretcher to the hydraulic piston and the stretcher tilting mechanism becoming loose.",
      "recall_initiation_date": "20151221",
      "center_classification_date": "20160413",
      "termination_date": "20180625",
      "report_date": "20160420",
      "code_info": "All Concerto & Basic Shower Trolleys manufactured by ArjoHuntleigh Polska Sp. z o.o, March 7, 2014 through April 20, 2015 with serial numbers from P0227559 - P0322393",
      "more_code_info": ""
    }
  ]
}