      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76374",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and internationally to  Argentina, Australia, Austria, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Canada, Chile, China, Columbia, Croatia, Czech Republic, Ecuador, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macau, Malaysia, Mexico, Moldova, Montenegro, Netherlands, New Zealand, Nigeria, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Republic and Venezuela.",
      "recall_number": "Z-1384-2017",
      "product_description": "iChemVELOCITY Urine Chemistry System, Catalog No. 800-3530, 800-7167, 800-7719, 800-7720, 800-3564, 800-7100, 800-7103, 800-7106, 800-7162, 800-7163, 800-7166, 800-3061.  The iChem Velocity automated urine chemistry system is used to test for the following analytes in urine: ketones, blood, protein, glucose, bilirubin, urobilinogen, nitrite, leukocyte esterase, specific gravity, pH, ascorbic acid, color, and clarity. The iChem Velocity automated urine chemistry analyzer is intended to be used for in vitro measurement of various analytes in urine; it is intended for use only with the iChem Velocity urine chemistry strips",
      "product_quantity": "3175 units total (1697 units in US)",
      "reason_for_recall": "The probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.",
      "recall_initiation_date": "20170130",
      "center_classification_date": "20170301",
      "termination_date": "20170819",
      "report_date": "20170308",
      "code_info": "Serial Number  V01257  V02416  V02619  V01444  V02906  V02959  V03934, V03933  V02027, V02022, V04322  V03288  V03314  V02330  V02326  V02329  V02642  V02323  V03571  V03428  V02909, V01251  V01115  V02081  V02248  V03575  V03854  V03403  V04387  V04388, V01406  V01556  V04328  V04390, V01557  V04141  V02103  V01214  V04227, V02438  v01838  V04284  V02535  V02534  V02449  V04312  V02538  V02635  V02530  V02450  V02634  V02485  V02436  V02614  V02463  V02652  V02502  V02536  V02464  V02423  V02435  V02432  V02649  V02459  V02537  V03883  V01689  V03409, V03413  V01753, V01754  V02243  V02985  V02242  V03976  V03851  V03925  V04369  V02026  V03848  V01404  V02430  V03565  V02234  V02373  V01829  V02446  V02703  V01429  V02949  V02899  V04309  V01318  V01546  V04257  V03566  V01548  V01547, V02753  V01541  V01542  V02369, V03922  V01555  V02181  V02070  V01762  V04357  V04350  V04349  V04356  V04291  V01720  V03834  V02335  V03798  V01911  V01998  V03567  V04258  V03238  V04329  V03327  V01776  V01712  V03271  V01966  V01824, V02086  V01467  V03100  V01872  V04138  V02047  V02210  V03221  V03003  V03285, V03283  V02025  V01179  V01182  V03033  V02361  V01842  V02931  V01717, V04377  V03776  V01831  V03890  V01335  V01567,V01568  V03237  V03000  V02998  V03464  V02046  V01581  V01493  V03842  V04379  V01460  V04348  V03113  V01146  V01781  V04302  V01558  V04195  V03251  V04153  V02650  V03417  V02935  V02185  V01432  V02150  V03456  V02782  V02194  V02078  V02127  V02217  V02810  V03748  V01533  V02957  V01864  V04385, V01632, V01633  V04318  V03008  V03011, V03047  V02492  V01135  V01161  V02602  V02934  V01482  V02186  V02884  V02411  V01200  V03245  V01627  V01877  V01094  V01092  V01303  V01332  V02962  V03090  V04341  V03977, V03953  V01876  V01147  V03797  V02809  V03467  V03700  V03775  V04335, V03357  V02180  V01376  V01978, V01981  V01985  V03438  V01486  V02460  V01642  V01616  V03427  V01502  V03889  V03105, V03106, V03107  V02269  V01713  V03384  V03716  V02808  V01620  V03858  V01649  V02295, V02306  V02468  V01059  V04372  V01645  V03441, V03376  V02874  V03145  V01375  V03853  V04242  V01919  V01236  V03187  V03698  V03825  V01389  V02398  V02129  V03382  V03055  V02195  V02201  V02097  V02093  V02085  V03470  V01326  V01237  V01505  V03920  V02018  V02706  V01834  V01400, V01401  V03573  V03836  V03388  V02656  V02674  V02414  V02314  V02456  V04311  V02861  V01641  V03402  V02419  V02558  V01243  V01244  V02239  V01973  V01612  V03474  V01442  V01388  V02429  V01294  V01298  V02069, V01159  V01393  V04279  V03383  V01730  V01046  V01274  V02717  V03958, V01078, V03957  V01117, V01118  V01752  V02517  V02587  V02868  V04313, V02920  V02965, V02964  V02130  V03439  V01534  V01664  V02505  V01250  V04375  V03576  V04382  V01770  V01377  V01431  V01072  V01862  V02317  V02287  V01134  V03186  V02367  V03796  V02453  V04236  V01977  V04256  V03114  V01360  V02881  V01553, V04391  V02796  V02958  V02852  V03819, V03827  V04278  V01309  V01948  V04248  V03423  V01314  V01076  V02992  V01893, V01889  V01361  V03434  V03419  V01459  V01915  V03133  V01437  V03463  V03481, V04193  V03163  V04345  V03335  V02875  V02193  V01871  V01755  V03046  V03978  V03086  V03089  V03088  V03087  V03085  V02988, V02983, V02995  V01599  V02756  V02123  V01625  V02568  V03444, V03444  V01929  V03859  V04353  V03143  V01149  V03372  V02067  V01184  V04097  V01610  V04401  V03864  V01329  V01281  V02697  V02727  V02065  V04367  V02509  V03849  V01132  V02054, V02033  V02036  V03560  V01540  V02849  V02724  V02646, V02665  V02462  V02020, V02021, V02966  V03318  V01551  V01989  V03356  V02529  V01313  V02956, V02952  V03860  V04117  V03793, V03794  V04137  V02110  V01788  V03303, V01170  V03254  V01765  V03274, V03275  V03316, V03331  V03278  V03201  V03043  V02669  V01213  V03813  V01570  V02190  V03440  V03082  V01874  V01210  V01462  V03918  V03786  V03422  V02144  V03057  V02337  V03415  V01990  V03370  V03391  V03915, V03250  V03281  V02679, V01077  V01594  V04320  V04274  V01901  V03379  V03790  V02087  V02501  V04308  V02664  V02327  V02491  V02448  V01138  V02451  V03040, V03386  V01173  V02951  V04340  V03167  V01461  V03079  V03702  V04319  V04238  V04381  V02457  V02452  V01183  V01421  V02219  V02955  V01936, V01934  V01941  V01945  V02551  V03914  V01958  V02147  V01980  V04241  V04339, V01572, V01571, V01566  V03815  V03803, V03824  V03888  V01160  V01769, V04380  V02143  V03425  V02854  V02172, V02171  V02023, V02009  V02383, V02384  V02255  V02249, V02254  V02380, V02381  V02019  V02259, V02263  V02233  V01942  V02298, V02299  V01938  V02252, V02256  V02409  V02251  V02257, V02289, V02284  V04310  V02285, V01852  V02253  V04270, V04272  V03781, V03780  V04246, V04247  V04142, V04143  V02051, V01996  V04240, V04239  V04263, V04264  V04223, V04224  V04266  V01853  V03812, V03804  V02003  V02250  V02286  V02258  V02729  V02146, V02188  V02288, V02322  V02138, V02099  V02245, V02231  V02170, V02091  V02139, V02140  V04292, V04253  V03871, V03872  V03749  V03779  V02961  V02313  V03955  V04362, V04360  V03111  V02533  V01686  V01304  V01305  V01545  V02789  V01430  V03874  V03758  V03473, V03473  V02892  V01169  V02963  V02948  V02164  V04093  V03866  V03049  V02167, V02014, V03919  V04288  V01857, V01856  V02082, V01233  V01225  V02984  V01198  V02897  V02169, V02412  V02158  V01719  V02166  V04286  V02117  V01714  V01816  V01666  V03102  V01850  V02910  V01903  V01296, V03852  V01563  V01957  V03387  V03817  V01564  V03380  V03828  V02017, V02004  V02310  V01395  V03058  V01061  V02685  V02893  V03433  V02699  V01293  V02580  V04267  V03774  V02267  V03935  V03401, V03451  V01386  V02319  V02971  V04249, V01423  V01586  V01951  V01544  V02620  V01935  V01255, V01199  V01413  V02102  V03932  V02131  V01196  V03373  V03258  V01967, V02470  V03189  V03156  V03833  V04262, V04327, V04407  V03723  V02010, V01994  V01367  V01858  V04336  V02541  V01205  V04295  V04294  V02031  V01405  V03830  V01055  V02497  V01075  V01639  V02630  V04255  V01409  V01107  V01463  V01789  V01279  V01315  V01192  V01849  V03877  V01427  V01252  V04307  V04298, V01552  V02356  V03279  v02901  V01316  V02015  V03929, V03928  V01325  V01759  V02226  V01081  V03364, V01387  V03891  V02151  V03101  V04354  V04252, V01122  V03421  V03400  V03855  V02359  V01189  V01176  V03130  V02608  V02539  V03759  V02723  V02725  V02275  V01615  V01333  V01726  V01909  V04358  V03367, V01439  V02886, V02859  V02804  V03699  V03975  V03841  V02503  V03777  V01940  V03351  V01504  V02421, V01501  V01503  V02354  V03476  V02094  V03282, V03273  V02066  V03466  V01477  V04250  V02472  V04323  V01954  V01870  V01841  V01410, V04304  V01525  V01334  V01778  V02903, V02904  V02115, V02114  V02308  V01324  V01786, V01785  V02837  V02247  V04096  V02107  V02454, V01216  V03411  V01565  V02098  V04359  V02927  V01270  V03873  V03381  V04409  V01905  V03450  V03458  V01402  V03244  V04331  V04261  V01249  V03108  V01623  V01509  V02960  V02136  V03767  V03153  V01223  V02715  V03365, V01306  V03369, V03358  V03371, V01307  V04111  V03954  V02035  V02783  V02494  V04101  V02519  V04393  V02527  V03110  V01590  V01411  V02191  V01239  V02896  V02581  V01716  V03176  V01101  V03078  V04343, V01767  V03483  V03329  V03472  V04265  V03432  V03437  V04334  V01640  V03856  V01913  V03454  V03442  V01282  V03249  V03725, V01636  V03901  V02272  V02476  V02274  V02888  V03720  V02396  V03009  V04276  V02202  V04254  V04098, V04099  V03561  V02836  V02944  V02490  V04363, V04364, V01820, V04365, V01819  V02154  V03276  V02894  V01038  V01129  V01465  V02063  V02100  V02096  V02428  V03460, V03416  V02126  V01576  V01037  V02071  V04140  V04321  V02680  V02235  V01104, V04368  V02716  V03921  V01440  V01403  V02673  V02420, V02424  V01190  V01860  V01308  V02973  V04114, V04115  V04273  V02933  V03926  V04280  V01492  V02801  V04251  V01904  V02092, V02134  V01569  V02954  V01722, V01130  V03426  V04112  V01772  V02426  V02161, V02176  V03142  V01655  V02375  V02731  V03385  V02830, V01227  V02028  V02628  V02638, V02124  V02996  V02552  V01180  V03475  V01619  V01869  V01163  v03791  V02858  V02336  V01475  V03001  V01271  V02877  V01372, V01736,V02883  V04116  V02880  V02016  V04305  V02867  V01235  V02179  V02183  V02427  V04285  V02182  V04230  V04231  V04232  V04228  V04234  V04293  V04283  V02244, V01956  V03412  V02583  V04118  V03876  V02946  V02709  V01507  V01297  V01476  V03746  V02812  V03835, V03857  V03290  V02072  V01601  V01832  V01771  V03304  V01984  V01201  V01208  V03765  V03377  V04094  V03206  V01768  V03161  V02212  V03200  V03917  V03850  V03816  V03199, V03198  V03846  V03875  V01744, V04303  V01892  V01099  V02891  V03076  V01095  V03139  V01757  V02895  V01261  V03722  V02189  V04347  V03019  V01855  V02149  V04376  V04271, V04314  V01584  V03745  V04399  V02080  V01843, V01818  V02758  V02331  V03005  V01529  V04229  V01321  V01847  V02216  V03004  V02197  V02923  V02749  V01186  V03072  V03116  V03399  V01217  V02898  V01425  V01397  V02467  V04297  V03715, V03436  V02208  V04301  V02056  V01873  V03792  V03789  V01212  V03773  V02402  V01782  V01979  V02012  V01330  V02878  V04289, V04290  V03443, V03443  V03952  V02350  V02885  V03568  V04237, V04219  V01999  V03788  V02601  V02349  V02876  V01878  V01550  V01970  V01178  V02073  V02152  V03461  V02013, V04408  V04398, V04317  V01562  V01103  V02168  V03353  V03270  V01707  V04384, V04383  V02076  V01614  V01299  V02192  V02422  V02767  V04113  V02871  V04135  V03719  V03923  V03916  V02024  V02924, V02921, V02922  V03404  V01535  V02187  V02431  V01142  V03930  V01727  V02211  V03900  V01371  V02547  V02553  V03084  V03471  V03482  V03144  V01141  V03410  V04386  V01394  V03447  V03562  V03563  V01932  V03418  V03564  V03453  V02826  V03847  V01156  V03264  V01780  V03902  V03724  V04299, V02559  V01215  V02390  V02722  V02929  V02104  V04222  V01229  V02643  V01224  V03477  V02887  V01428  V03870  V03420  V01588  V02207  V04134  V02095  V02526  V02074  V01613  V03269  V01148  V02882  V02122, V03956  V01287  V01916  V01531  V01595  V03865  V01827  V04275  V04277  V01952  V01331  V02406  V01131, V01133  V01888  V02866  V01424  V02710  V02682  V02403  V01319  V02626  V03863, V02970  V03801  V01987  V01317  V02765, V02761  V03768  v01837  V03136  V02418, V02413  V01803  V03128  V03747  V02521  V01166  V01512  V01914, V01436  V01450  V01920, V01918  V03924  V01763  V01805  V03480  V01737  V01825  V04306  V02543  V04315  V01426  V03032, V03026  V01506  V02371, V02368  V01204, V01203  V01154  V01798  V01119  V03366  V02730  V02342  V01775  V01202  V01491  V03867  V01479  V01634  V01471  V04226, V01517  V01734  V02052  V01766  V01125  V02006, V02007  V01100  V01516  V01743  V01167  V01074, V02914  V01543  V02748  V04139  V01123  V02676  V01466  V01804  V03355  V02005  V03138, V03141  V01522  V01449  V01575  V03408  V04269, V04268  V03219  V02582, V02566  V04351, V01583, V04326  V02294  V04110, V04109  V01600  V01218  V03123, V02556  V02757  V01483  V01519, V04235  V01470  V02407  V03308  V03785  V03892  V01058  V01513  V03405, V03569  V02705  V02805  V01064  V01358  V03869  V03829  V01986, V01983  V02743  V02586  V02162  V01523  V04225  V02318  V02813  V01080, V01086  V01602  V01181  V02991, V01054  V01965  V04416, V04392, V01573  V02869  V01521  V04361, V04404  V01962  V01526, V04370  V03868  V01733  V01611  V02763  V01073, V01048  V01669  V01706, V01705  V01696, V01695  V01702, V01704  V02513  V01191  V01480  V01070  V01764  V04316, V01382  V03051, V02057  V01097  V02292  V01598  V02215  V01665, V01761  V02029, V03766  V01108  V03744  V02942  V01469  V04411, V04378  V01042  V04402  V01082  V04412  V02372  V01731  V01750  V01708  V01693, V02712  V01043  V01884  V03081, V01044  V02764  V01597, V01593  V01683  V02382  V01057  V01779  V03309  V01398, V04095  V01084, V01056  V03837, V01668  V04330, V02309  V01248, V02037  V01511  V01760  V01674  V01518, V04244  V02793  V04405  V02773  V01582, V04406  V01530, V04233  V03814, V01520  V03974, V03750  V01701, V01700  V02644  V03006  V01083, V01008  V02737  V01185  V01320  V01964  V01228, V02890  V02042, V04136  V01963  V03302  V01532, V04355  V01484, V01485  V01071  V02675  V03732  V02040  V01277  V01481  V02206, V02120  V01718  V01194, V01193  V01738  V01729  V02915, V04400, V01381  V01673, V04300  V03886, V01670  V01524  V04337, V02466, V04338  V03802  V04102  V01416  V02508  V01399  V03253  V02755  V03484, V04366  V01145  V03406, V03799  V03818  V03826  V02125  V01254  V04403  V02862  V03927  V02240  V02821  V02209, V03884  V02227  V02982  V02061, V02055  V04324  V02651, V02647  V03574, V01197  V02846, V02845  V02733  V02916, V01165  V03435, V03701  V01561  V01310, V01311  V03731  V02109  V03730  V01875, V01844  V01242, V01240  V04373  V04296, V03903  V03424  V01174  V03469  V02918, V02932  V03931  V02374  V01728  V02048  V03132  V01323  V02106  V01284  V03787  V02512  V02204  V01746, V01747  V03479  V01756  V01356  V03478  V02077  V03893  V03080  V04325  V03714  V01438  V03717, V03718  V03721  V01637, V01638  V04346, V02377  V01003  V03109  V03104  V02567  V02068  V02079  V02089  V02986  V02334  V04352  V02855  V04344  V01222  V03784  V01087  V01749, V01245  V04260  V02711  V01643  V04371  V03048  V01538  V03795  V03894  V04389, V01559  V03272  V01732  V02999  V03885  V01458  V02064  V03800  V04342  V01096  V02721  V03448  V03445  V03449  V03446  V01036  V02560, V02555  V02393, V02395  V01412,V03570   V03012  V02307  V03468  V04374  V01188  V01238  V01447, V01446  V04100  V02829, V02873, V02863  V03124  V02870, V01384  V03491, V03547  V02702  V02304  V03301  V02629  V03042  V02198, V02199, V03609  V02905  V03965  V03610  V03341, V02814, V02775, V03344, V02772  V02800  V03712, V02610, V02698, V02632  V02786, V02774, V03277, V02787, V03340  V02781  V02564  V02850  V03266, V03263  V03015  V02947  V03041  V02425  V02362  V03056  V04021",
      "more_code_info": ""
    }
  ]
}