{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87475",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland,   France, Germany, Hong Kong, Ireland, Italy, Netherlands,   Norway, Poland, Portugal, Spain, Sweden, Switzerland,   United Kingdom.",
      "recall_number": "Z-1383-2021",
      "product_description": "Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long  Item Code: SIGSDL45CTVT",
      "product_quantity": "5928 units WW: 4212 US  1716 OUS",
      "reason_for_recall": "Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.",
      "recall_initiation_date": "20210304",
      "center_classification_date": "20210409",
      "termination_date": "20240712",
      "report_date": "20210421",
      "code_info": "Lot Numbers:  N0A1038UY  N0E0016UY  N0F0616UY  N0K0934UY  N0B0493UY  N0F0254UY  N0F0817UY  N0M0334UY  N0C0436UY  N0F0518UY  N0G0388UY  N9K1095UY  N0D0731UY  N0F0615UY  N0J0859UY  N9L0797UY    UDI:  10884521741843  20884521741840"
    }
  ]
}