  "results": [
    {
      "status": "Ongoing",
      "city": "North Billerica",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90433",
      "recalling_firm": "Breas Medical, Inc.",
      "address_1": "16 Esquire Rd",
      "address_2": "",
      "postal_code": "01862-2527",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1382-2022",
      "product_description": "Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation    Model Number: 230000     ¿\tIf the recalled product is a software, include version number: firmware versions 5.0.5 or earlier",
      "product_quantity": "1,113 units",
      "reason_for_recall": "A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained",
      "recall_initiation_date": "20220602",
      "center_classification_date": "20220713",
      "report_date": "20220720",
      "code_info": "Unique Device Identifier (UDI): UDI 07321822300004.    Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9.",
      "more_code_info": ""
    }
  ]
}