{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bloomington",
      "address_1": "750 N Daniels Way",
      "reason_for_recall": "The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.",
      "address_2": "",
      "product_quantity": "483",
      "code_info": "All  lots",
      "center_classification_date": "20180417",
      "distribution_pattern": "Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)",
      "state": "IN",
      "product_description": "MINC Benchtop Incubator (Mini Incubator)  catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.",
      "report_date": "20180425",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cook Inc.",
      "recall_number": "Z-1377-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79711",
      "termination_date": "20200508",
      "more_code_info": "",
      "recall_initiation_date": "20180112",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}