      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65042",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-1376-2013",
      "product_description": "NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm.     Used to prepare cylindrical holes in the native tibia.",
      "product_quantity": "69 units",
      "reason_for_recall": "Zimmer is initiating a lot specific recall of the Trabecular Metal\" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the applicable drill guide.",
      "recall_initiation_date": "20130327",
      "center_classification_date": "20130524",
      "termination_date": "20170126",
      "report_date": "20130605",
      "code_info": "Part No. 00-5887-052-00; lots 62181290 and 62183430"
    }
  ]
}