{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90363",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AL\tAR\tAZ\tCA\tCO\tCT\tDC\tFL\tGA\tHI\tIA\tIL\tIN\tKS\tKY\tMA\tMD\tMI\tMN\tMO\tMS\tMT\tNC\tNE\tNJ\tNY\tOH\tOK\tOR\tPA\tPR\tSC\tSD\tTN\tTX\tUT\tVA\tWA\tWI\tWV    OUS: Canada, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Poland, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Argentina, Australia, Brazil, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Israel, Japan, Kuwait, Lebanon, Malaysia, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates (UAE), Vietnam",
      "recall_number": "Z-1374-2022",
      "product_description": "Ventana HE 600 System, automated slide preparer, for use in laboratories.",
      "product_quantity": "763 instruments",
      "reason_for_recall": "There is a potential for fluid leak  inside the Ventana HE 600 instrument which could result in electrical short and fire",
      "recall_initiation_date": "20220601",
      "center_classification_date": "20220712",
      "report_date": "20220720",
      "code_info": "All Lots; UDI/DI: 04015630976010",
      "more_code_info": ""
    }
  ]
}