  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87486",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA    O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.",
      "recall_number": "Z-1373-2021",
      "product_description": "MicroScan WalkAway-96 plus Reconditioned Instrument",
      "product_quantity": "9 systems",
      "reason_for_recall": "Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.",
      "recall_initiation_date": "20190904",
      "center_classification_date": "20210407",
      "termination_date": "20230111",
      "report_date": "20210414",
      "code_info": "Catalog Number: B1018-396R  UDI Code: 15099590684129  Serial Numbers:  3967245  3967417  3967493  3967544  3967634  3967734  3967914  3928224  3968442"
    }
  ]
}