  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87486",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA    O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.",
      "recall_number": "Z-1372-2021",
      "product_description": "MicroScan WalkAway-40 plus Reconditioned Instrument",
      "product_quantity": "6 systems",
      "reason_for_recall": "Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.",
      "recall_initiation_date": "20190904",
      "center_classification_date": "20210407",
      "termination_date": "20230111",
      "report_date": "20210414",
      "code_info": "Catalog Number: B1018-340R  UDI Code: 15099590684112  Serial Numbers:  34020035  34020224  34030054  3407179  3407459  3407532"
    }
  ]
}