  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87486",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA    O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.",
      "recall_number": "Z-1370-2021",
      "product_description": "MicroScan WalkAway-40 plus Instrument",
      "product_quantity": "16 systems",
      "reason_for_recall": "Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.",
      "recall_initiation_date": "20190904",
      "center_classification_date": "20210407",
      "termination_date": "20230111",
      "report_date": "20210414",
      "code_info": "Catalog Number: B1018-283  UDI Code: 1509959065866  Serial Numbers:  34010001  34010002  34010003  34010004  34010005  34010006  34010007  34010008  34010009  34010010  34010011  34010012  34010013  34010014  34010015  34010016"
    }
  ]
}