  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70678",
      "recalling_firm": "BD Biosciences, Systems & Reagents",
      "address_1": "2350 Qume Dr",
      "address_2": "N/A",
      "postal_code": "95131-1812",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim",
      "recall_number": "Z-1370-2015",
      "product_description": "BD FACSCalibur;  Catalog Numbers 342973, 342975, 34976    In vitro Diagnostics Use.",
      "product_quantity": "72 devices",
      "reason_for_recall": "The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.",
      "recall_initiation_date": "20150306",
      "center_classification_date": "20150402",
      "termination_date": "20160404",
      "report_date": "20150408",
      "code_info": "Serial Nos.  E34297300656,  E34297300657, E34297300659, E34297300660, E34297300661,  E34297300666, E34297300667, E34297300668, E34297300669,  E34297300670, E34297300671, E34297300673,  E34297300675, E34297300676, E34297300677,   E34297300678    E34297502563, E34297502564,  E34297502565, E34297502566, E34297502567,  E34297502568, E34297502569, E34297502570, E34297502571,  E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578,  E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586,  E34297502587, E34297502588, E34297502589,  E34297502590, E34297502591, E34297502592, E34297502593,  E34297502594, E34297502595, E34297502596,  E34297502597, E34297502599, E34297502600, E34297502601,  E34297502602, E34297502603, E34297502604,  E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619,  E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502,  E34297600503, E34297600507, E34297600508."
    }
  ]
}