{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Beach Gardens",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84771",
      "recalling_firm": "Biomet 3i, LLC",
      "address_1": "4555 Riverside Dr",
      "address_2": "N/A",
      "postal_code": "33410-4200",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:  Australia  Austria  Belgium  China  France  Germany  Italy  Japan  Netherlands  Poland  Spain  UK",
      "recall_number": "Z-1369-2020",
      "product_description": "CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 5MM(P) X 3MM(H)  Reference Number: IEHA353",
      "product_quantity": "N/A",
      "reason_for_recall": "Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin contour misalignment",
      "recall_initiation_date": "20200110",
      "center_classification_date": "20200226",
      "termination_date": "20210802",
      "report_date": "20200304",
      "code_info": "Lot Numbers: 1228667"
    }
  ]
}