      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70775",
      "recalling_firm": "Roche Diabetes Care, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa.  The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.",
      "recall_number": "Z-1369-2015",
      "product_description": "ACCU-CHEK Connect Diabetes Management App; Instruction Manual    Designed to transfer data for diabetes management.",
      "product_quantity": "The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.",
      "reason_for_recall": "Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user.  A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb",
      "recall_initiation_date": "20141030",
      "center_classification_date": "20150402",
      "termination_date": "20150608",
      "report_date": "20150408",
      "code_info": "N/A"
    }
  ]
}