{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64850",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "USA Nationwide Distribution",
      "recall_number": "Z-1368-2013",
      "product_description": "Cobas¿ 8000 modular analyzer series software version 03-01, 02, 03 and 04, impacting the cobas e 602 module only.    Product Usage:  Fully automated immunoassay analyzer intended for the in-vitro quantitative/qualitative determination of analytes in body fluids. This analyzer is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes and photometric analysis modules",
      "product_quantity": "72 units",
      "reason_for_recall": "When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met simultaneously: More than a total of 2000 reagent kits or diluents have been registered on the cobas e 602 modules within one system core, and two different diluent",
      "recall_initiation_date": "20130404",
      "center_classification_date": "20130523",
      "termination_date": "20150224",
      "report_date": "20130529",
      "code_info": "Part Number 05990378001  Serial Numbers 1017-06, 1021-03, 1021-05, 1142-09, 1145-10, 1140-01, 1125-07, 1128-02, 1146-01, 1146-02, 1146-03, 1171-10, 1169-06, 1169-08, 1169-09, 1172-09, 1156-01, 1156-02, 1156-09, 1160-04, 1160-06, 1016-05, 1017-01, 1139-08, 1139-10, 1140-07, 1140-08, 1140-10, 1156-06, 1156-07, 1158-09, 1160-01, 1160-05, 1173-01, 1139-06, 1139-07, 2374-11, 1127-08, 1138-08, 1138-10, 1139-01, 1156-08, 1156-10, 1173-10, 1156-05, 1158-10, 1139-09, 1131-08, 1132-01, 1132-02, 1171-07, 1171-08, 1171-09, 1147-05, 1149-01, 1124-05, 1124-06, 1132-03, 1132-04, 1147-05, 1149-01, 1128-05, 1128-06, 1139-02, 1139-03, 1139-04, 1139-05, 1146-05, 1146-06, 1147-04, 1156-04, 1132-05, 1132-10, 1137-02, 1147-05, 1149-01, 1128-05, 1128-06, 1149-07, 1149-08, and 1167-02."
    }
  ]
}