      "status": "Ongoing",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90399",
      "recalling_firm": "Wright Medical Technology, Inc.",
      "address_1": "1023 Cherry Rd",
      "address_2": "",
      "postal_code": "38117-5423",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico",
      "recall_number": "Z-1366-2022",
      "product_description": "Wright EVOLVE TRIAD Plate Cutter, REF 49510120",
      "product_quantity": "231",
      "reason_for_recall": "The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use",
      "recall_initiation_date": "20220614",
      "center_classification_date": "20220711",
      "report_date": "20220720",
      "code_info": "Lot codes:  955401, 1168853, 1349302, 1956936, 2486857",
      "more_code_info": ""
    }
  ]
}