  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "61315",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide including the state of  VA and the country of Ireland.",
      "recall_number": "Z-1366-2013",
      "product_description": "Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452    Product Usage:  An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.",
      "product_quantity": "20 liners",
      "reason_for_recall": "DePuy Orthopaedics, Inc.  (DePuy) is issuing a voluntary recall of one lot of the AltrX\" liner because the product was incorrectly labeled on the box.  The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product enclosed was a 36IDX56OD Neutral Liner.  Three ( 3 ) complaints were received in January 2012.",
      "recall_initiation_date": "20120307",
      "center_classification_date": "20130522",
      "termination_date": "20130524",
      "report_date": "20130529",
      "code_info": "product code1221-36-452,  lot 187460."
    }
  ]
}