  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90423",
      "recalling_firm": "Draeger Medical Systems, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "No US distribution. Foreign distribution to BY, CL, DE, ES, GB, HU, IN, IT, KW, MX, SA, SV, TW.",
      "recall_number": "Z-1365-2022",
      "product_description": "Draeger Globe-Trotter and Globe-Trotter IC, Model numbers MS20508 and MS20509. neonatal transport system",
      "product_quantity": "38 OUS (no US distribution)",
      "reason_for_recall": "The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.",
      "recall_initiation_date": "20220520",
      "center_classification_date": "20220711",
      "report_date": "20220720",
      "code_info": "Model, country, and serial number:    MS20508:  El Salvador DSPN-0003   Mexico DSPM-0003, DSPH-0002  Chile DSPL-0003  Saudi Arabia DSPJ-0003, DSPL-0004, DSPL-0005, DSPL-0006, DSPL-0007, DSPM-0001, DSPM-0002, DSPL-0008, DSPK-0002, DSPH-0003, DSPJ-0004, DSRA-0001  Kuwait DSPJ-0002, DSPK-0001  India DSPN-0001, DSPN-0002, DSRA-0002  Taiwan DSPL-0001, SPL-0002, DSPJ-0001    MS20509:  Germany DSPH-0003  Qatar DSPL-0007, DSRB-0001  Spain DSPK-0002, DSPJ-0006, DSPJ-0003, DSPH-0001, DSPJ-0004, DSPK-0001, DSPL-0008  Hungary DSPF-0003  Belarus DSPJ-0005  United Kingdom DSPF-0002  Italy DSPJ-0002",
      "more_code_info": ""
    }
  ]
}