  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Grenoble Cedex 1",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84766",
      "recalling_firm": "BIOMERIEUX - Grenoble",
      "address_1": "Biomerieux",
      "address_2": "Institut Christophe Merieux; 5 Rue Des Berges; Cs 30076",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "No US distribution. No Mexican consignees. International distribution to South Africa, Uruguay, Tanzania, Romania, Netherlands, Macao, Italy, India, Hong Kong, United Kingdom, France, Spain, Germany, China, Chile, Canada, Belgium, Australia, Argentina, Curacao.",
      "recall_number": "Z-1365-2020",
      "product_description": "NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS  Manufacturer  bioMerieux SA  376, Chemin de l'Orme  69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.",
      "product_quantity": "1024",
      "reason_for_recall": "Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201.  The issue is linked to drift of pH electrode dysfunction.",
      "recall_initiation_date": "20200110",
      "center_classification_date": "20200226",
      "termination_date": "20220414",
      "report_date": "20200304",
      "code_info": "Lot 19040201, Expiry 28-Feb-2021"
    }
  ]
}