{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90411",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St Bld 950",
      "address_2": "",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Nationwide distribution in the states of NC and WV.",
      "recall_number": "Z-1364-2022",
      "product_description": "AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)",
      "product_quantity": "2 units",
      "reason_for_recall": "AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation",
      "recall_initiation_date": "20210803",
      "center_classification_date": "20220711",
      "termination_date": "20240410",
      "report_date": "20220720",
      "code_info": "UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502",
      "more_code_info": ""
    }
  ]
}