  "results": [
    {
      "status": "Ongoing",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84961",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CO and FL.",
      "recall_number": "Z-1364-2020",
      "product_description": "Truliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in primary and revision total knee arthroplasty.",
      "product_quantity": "4 units",
      "reason_for_recall": "These Truliant Splined Stem Extension, 18mm x 120mm were found to have a labeling error. The upper right corner of the label incorrectly identifies the implant size as  17mm x 80mm  when the implant size is 18mm x 120mm.",
      "recall_initiation_date": "20200127",
      "center_classification_date": "20200226",
      "report_date": "20200304",
      "code_info": "Catalog no: 02-012-64-1812    Serial number: 6366405, 6367075, 6367076"
    }
  ]
}