{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Burmoos",
      "state": "",
      "country": "Austria",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79533",
      "recalling_firm": "W & H DentalWerk Burmoos GMBH - Site 1",
      "address_1": "Ignaz-Glaser-Strasse 53",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1351-2018",
      "product_description": "Foot Controls  Model: S-N1 and S-N2    Product Usage:  Variable foot control for use with drive units operating mechanical instruments.",
      "product_quantity": "",
      "reason_for_recall": "During continuous postmarket surveillance activities, we have noticed a potential malfunction of individual foot controls of the type S-N1/S-N2.  In case the error occurs, the foot control might not switch back to zero position when the pedal is released. Consequently, the motor keeps rotating. If this happens during automatic thread cutting or automatic screwing-in of implants, it may, in the worst case, lead to a serious injury of the patient.",
      "recall_initiation_date": "20171221",
      "center_classification_date": "20180410",
      "termination_date": "20190607",
      "report_date": "20180418",
      "code_info": "REF:  05046200    05046200    05046211    05046210    06202400    06202400    30285000   30285001     Serial number \"SN\"  from  to (including)  22400 23429  E02201 E02485  E01011 E01033  E01035 E01071  24315 34040  E01358 E01658  01001 02318",
      "more_code_info": ""
    }
  ]
}