{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Harleysville",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70472",
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "address_1": "1499 Delp Dr",
      "address_2": "N/A",
      "postal_code": "19438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-1351-2015",
      "product_description": "Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III;    AngioDynamics product- Dura-Flow Catheter, Schon XL Catheter; The affected product are insertion kits used to implant the devices packaged-long and short term hemodialysis catheters and infusion catheters.",
      "product_quantity": "3152",
      "reason_for_recall": "Medcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI (biological indicator) out of 21 on the sterilization load was out of specification.",
      "recall_initiation_date": "20150205",
      "center_classification_date": "20150330",
      "termination_date": "20150806",
      "report_date": "20150408",
      "code_info": "Medcomp products:Catalog Code/ Lot Number/Expiration Date:  DFXL148MTE/ MGMC610/10/31/2019; HFS 32/ MGMB440/ 07/14/2017;   HFS24E./MGLY830/07/31/2017; MC061402/MGMD190/07/15/2017;   MC3L-8S/MGMD920/10/16/2019; MR17035211/MGMD680/10/31/2017;   TRAY #593-3/MGMB460/07/15/2017; RMS23602/MBWM930/07/31/2017;   ASPC2816-3/MBWM900/07/31/2017; Tray #55216-3/MBWL780/07/31/2017;  ASPC2816-3PC/MBWX290/07/31/2017; 10301207/MGMD930/07/31/2017;  10800701/ MGLY730/ 07/31/2017; 10800702/ MGMD210/ 07/31/2017;  10800703/MGLY740/07/31/2017      AngioDynamics products- model numbers/lot numbers/ expiration dates:    10301207/MGMD930/07/31/2017; 10800701/MGLY730/07/31/2017;  10800702/MGMD210/07/31/2017; 10800703/MGLY740/07/31/2017"
    }
  ]
}