{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Aliso Viejo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87353",
      "recalling_firm": "Nuvasive Specialized Orthopedics Inc",
      "address_1": "101 Enterprise Ste 100",
      "address_2": "N/A",
      "postal_code": "92656-2604",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution, including U.S. Nationwide.",
      "recall_number": "Z-1350-2021",
      "product_description": "NUVASIVE PRECICE UNYTE System (Humerus and Tibia):    PRECICE Trauma Nail System/PRECICE Unyte-Humerus  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 165 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 180 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 195 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 210 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 225 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 240 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 255 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 270 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 285 mm length;  Orthopedic trauma intramedullary rod, humerus, 8.5 mm diameter, 300 mm length;  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 0mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 2mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 5mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 10mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 15mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 0mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 2mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 5mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 10mm);  End Cap for PRECICE Trauma Intramedullary Nail System Implant (humeral nail, 15mm);  3.5 mm Locking Screw, 20 mm Length;  3.5 mm Locking Screw 25 mm Length;  3.5 mm Locking Screw 30 mm Length;  3.5 mm Locking Screw 35 mm Length;  3.5 mm Locking Screw 40 mm Length;  3.5 mm Locking Screw 45 mm Length;  3.5 mm Locking Screw 50 mm Length;  3.5 mm Locking Screw 55 mm Length;  3.5 mm Locking Screw 60 mm Length;  Locking Screw, 3.5x65mm;  Locking Screw, 3.5x70mm;  Locking Screw, 3.5x75mm;  5.0 mm Locking Screw 20 mm Length;  5.0 mm Locking Screw 25 mm Length;  5.0 mm Locking Screw 30 mm Length;  5.0 mm Locking Screw 35 mm Length;  5.0 mm Locking Screw 40 mm Length;  5.0 mm Locking Screw 45 mm Length;  5.0 mm Locking Screw 50 mm Length;  5.0 mm Locking Screw 55 mm Length;  5.0 mm Locking Screw 60 mm Length;  5.0 mm Locking Screw 65 mm Length;  5.0 mm Locking Screw 70 mm Length;  5.0 mm Locking Screw 75 mm Length;  5.0 mm Locking Screw 20 mm Length;  5.0 mm Locking Screw 25 mm Length;  5.0 mm Locking Screw 30 mm Length;  5.0 mm Locking Screw 35 mm Length;  5.0 mm Locking Screw 40 mm Length;  5.0 mm Locking Screw 45 mm Length;  5.0 mm Locking Screw 50 mm Length;  5.0 mm Locking Screw 55 mm Length;  5.0 mm Locking Screw 60 mm Length;  5.0 mm Locking Screw 65 mm Length;  5.0 mm Locking Screw 70 mm Length;  5.0 mm Locking Screw 75 mm Length;  5.0 mm X 80 mm Locking Screw, Peg;   5.0 mm X 85 mm Locking Screw, Peg;  5.0 mm X 90 mm Locking Screw, Peg;   5.0 mm X 95 mm Locking Screw, Peg;   5.0 mm X 100 mm Locking Screw, Peg;    PRECICE Trauma Nail System/Precice Unyte-Tibia  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 240 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 255 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 270 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 285 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 300 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 315 mm length;  Orthopedic trauma intramedullary rod, antegrade tibia 10 degree bend, 11.5 mm diameter, 330 mm length;  Orthopedic trauma intra",
      "product_quantity": "229 units",
      "reason_for_recall": "Due to not having the full complement of biological assessments for all potential patients and as a result the devices are not indicated for use in individuals under the age of 18 years old.",
      "recall_initiation_date": "20210212",
      "center_classification_date": "20210405",
      "report_date": "20210414",
      "code_info": "All Lots manufactured between April 2015 to February 2021.    PRECICE Unyte Product Codes (for Humerus):      Trauma intramedullary rod, humerus, 8.5 mm diameter:    CDH8.5-20L165;  CDH8.5-20L180;  CDH8.5-30L195;  CDH8.5-50L210;  CDH8.5-50M225;  CDH8.5-50M240;  CDH8.5-80M255;  CDH8.5-80M270;  CDH8.5-80M285;  CDH8.5-80M300;        end cap for trauma intramedullary humeral nail:    HEC1-000;  HEC2-000;  HEC3-000;  HEC4-000;  HEC5-000;        Locking screws:    LSB3-020;  LSB3-025;  LSB3-030;  LSB3-035;  LSB3-040;  LSB3-045;  LSB3-050;  LSB3-055;  LSB3-060;  LSB3-065;  LSB3-070;  LSB3-075;  LSB5-020;  LSB5-025;  LSB5-030;  LSB5-035;  LSB5-040;  LSB5-045;  LSB5-050;  LSB5-055;  LSB5-060;  LSB5-065;  LSB5-070;  LSB5-075;  LSC5-020;  LSC5-025;  LSC5-030;  LSC5-035;  LSC5-040;  LSC5-045;  LSC5-050;  LSC5-055;  LSC5-060;  LSC5-065;  LSC5-070;  LSC5-075;  LSC5-080;  LSC5-085;  LSC5-090;  LSC5-095;  LSC5-100.    GTIN Number (for Humerus):  00812258022730;  00812258022747;  00812258022754;  00812258022761;  00812258022778;  00812258022785;  00812258022792;  00812258022808;  00812258022815;  00812258022822;  00812258028121;  00812258028138;  00812258028145;  00812258028152;  00812258028169;  00812258021337;  00812258021344;  00812258021351;  00812258021368;  00812258021375;  00812258021382;  00812258021399;  00812258021405;  00812258021412;  00887517936233;  00887517936240;  00887517936257;  00856719002886;  00856719002893;  00856719002909;  00856719002916;  00856719002923;  00856719002930;  00856719002947;  00856719002954;  00856719002961;  00856719002978;  00856719002985;  00856719002992;  00812258021511;  00812258021528;  00812258021535;  00812258021542;  00812258021559;  00812258021566;  00812258021573;  00812258021580;  00812258021597;  00812258021603;  00812258021610;  00812258021627;  00887517001030;  00887517001047;  00887517001054;  00887517001061;  00887517001078;    Precice Unyte Product Codes (for Tibia):      trauma intramedullary rod, 11.5 mm diameter:    CDT11.5-30T240;  CDT11.5-30T255;  CDT11.5-30T270;  CDT11.5-30T285;  CDT11.5-30T300;  CDT11.5-30T315;  CDT11.5-30T330;  CDT11.5-30T345;  CDT11.5-30T360;  CDT11.5-30T375;  CDT11.5C-30T240;  CDT11.5C-30T255;  CDT11.5C-30T270;  CDT11.5C-30T285;  CDT11.5C-30T300;  CDT11.5C-30T315;  CDT11.5C-30T330;  CDT11.5C-30T345;  CDT11.5C-30T360;  CDT11.5C-30T375;      trauma intramedullary rod, 12.5 mm diameter:    CDT12.5-30T240;  CDT12.5-30T255;  CDT12.5-30T270;  CDT12.5-30T285;  CDT12.5-30T300;  CDT12.5-30T315;  CDT12.5-30T330;  CDT12.5-30T345;  CDT12.5-30T360;  CDT12.5-30T375;  CDT12.5C-30T240;  CDT12.5C-30T255;  CDT12.5C-30T270;  CDT12.5C-30T285;  CDT12.5C-30T300;  CDT12.5C-30T315;  CDT12.5C-30T330;  CDT12.5C-30T345;  CDT12.5C-30T360;  CDT12.5C-30T375;        end cap for trauma intramedullary nail:     TEC1-000;  TEC1-005;  TEC1-010;  TEC1-015;  TEC1-020;   Locking screws:      TSA4.5-020;  TSA4.5-025;  TSA4.5-030;  TSA4.5-035;  TSA4.5-040;  TSA4.5-045;  TSA4.5-050;  TSA4.5-055;  TSA4.5-060;  TSA4.5-065;  TSA4.5-070;  TSA4.5-075;  TSA4-020;  TSA4-022.5;  TSA4-025;  TSA4-027.5;  TSA4-030;  TSA4-032.5;  TSA4-035;  TSA4-037.5;  TSA4-040;  TSA4-042.5;  TSA4-045;  TSA4-047.5;  TSA4-050;  TSA4-055;  TSA4-060;  TSA4-065;  TSA4-070;  TSA4-075;  TSA4-080;  TSA4-085;  TSA4-090;  TSA4-095;  TSA4-100;  TSA4-100;  TSA5-020;  TSA5-020;  TSA5-022.5;  TSA5-022.5;  TSA5-025;  TSA5-025;  TSA5-027.5;  TSA5-027.5;  TSA5-030;  TSA5-030;  TSA5-032.5;  TSA5-032.5;  TSA5-035;  TSA5-035;  TSA5-037.5;  TSA5-037.5;  TSA5-040;  TSA5-040;  TSA5-042.5;  TSA5-042.5;  TSA5-045;  TSA5-045;  TSA5-047.5;  TSA5-047.5;  TSA5-050;  TSA5-050;  TSA5-052.5;  TSA5-055;  TSA5-055;  TSA5-057.5;  TSA5-060;  TSA5-060;  TSA5-062.5;  TSA5-065;  TSA5-065;  TSA5-067.5;  TSA5-070;  TSA5-070;  TSA5-072.5;  TSA5-075;  TSA5-075;  TSA5-080;  TSA5-080;  TSA5-085;  TSA5-085;  TSA5-090;  TSA5-090;  TSA5-095;  TSA5-095;  TSA5-100;  TSA5-100;  TSA5-105;  TSA5-105;  TSA5-110;  TSA5-110.       GTIN Number (for Tibia):  00812258027322;  00812258027339;  00812258027346;  00812258027353;  00812258027360;  00812258027377;  00812258027384;  00812258027391;  00812258027407;  00812258027414;  00812258027421;  00812258027438;  00812258027445;  00812258027452;  00812258027469;  00812258027476;  00812258027483;  00812258027490;  00812258027506;  00812258027513;  00812258027520;  00812258027537;  00812258027544;  00812258027551;  00812258027568;  00812258027575;  00812258027582;  00812258027599;  00812258027605;  00812258027612;  00812258027629;  00812258027636;  00812258027643;  00812258027650;  00812258027667;  00812258027674;  00812258027681;  00812258027698;  00812258027704;  00812258027711;  00812258028039;  00812258028046;  00812258028053;  00812258028060;  00812258028077;  00887517936011;  00887517936028;  00887517936035;  00887517936042;  00887517936059;  00887517936066;  00887517936073;  00887517936080;  00887517936097;  00887517936202;  00887517936219;  00887517936226;  00887517000033;  00887517000040;  00887517000057;  00887517000064;  00887517000071;  00887517000088;  00887517000095;  00887517000101;  00887517000118;  00887517000125;  00887517000132;  00887517000149;  00887517000156;  00887517000163;  00887517000170;  00887517000187;  00887517000194;  00887517000200;  00887517000217;  00887517000224;  00887517000231;  00887517000248;  00887517000255;  00887517000255;  00812258027728;  00887517000262;  00812258027735;  00887517000279;  00812258027742;  00887517000286;  00812258027759;  00887517000293;  00812258027766;  00887517000309;  00812258027773;  00887517000316;  00812258027780;  00887517000323;  00812258027797;  00887517000330;  00812258027803;  00887517000347;  00812258027810;  00887517000354;  00812258027827;  00887517000361;  00812258027834;  00887517000378;  00812258027841;  00887517000385;  00812258027858;  00812258027865;  00887517000392;  00812258027872;  00812258027889;  00887517000408;  00812258027896;  00812258027902;  00887517000415;  00812258027919;  00812258027926;  00887517000422;  00812258027933;  00812258027940;  00887517000439;  00812258027957;  00887517000446;  00812258027964;  00887517000453;  00812258027971;  00887517000460;  00812258027988;  00887517000477;  00812258027995;  00887517000484;  00812258028008;  00812258028008;  00812258028015;  00812258028015;    Precice Unyte Product Codes (covers both Humerus and Tibia):      trauma intramedullary rod, 10.7 mm diameter:    T10.7-50A205;  T10.7-50A225;  T10.7-50A240;  T10.7-50B205;  T10.7-50B225;  T10.7-50B240;  T10.7-50C205;  T10.7-50C225;  T10.7-50C240;  T10.7-50D205;  T10.7-50D225;  T10.7-50D240;  T10.7-50E205;  T10.7-50E225;  T10.7-50E240;  T10.7-80A255;  T10.7-80A285;  T10.7-80A315;  T10.7-80A345;  T10.7-80A375;  T10.7-80B255;  T10.7-80B285;  T10.7-80B315;  T10.7-80B345;  T10.7-80B375;  T10.7-80C255;  T10.7-80C285;  T10.7-80C315;  T10.7-80C345;  T10.7-80C375;  T10.7-80D255;  T10.7-80D285;  T10.7-80D315;  T10.7-80D345;  T10.7-80D375;  T10.7-80E255;  T10.7-80E285;  T10.7-80E315;  T10.7-80E345;  T10.7-80E375;      trauma intramedullary rod, 12.5 mm diameter:   T12.5-50A205;  T12.5-50A225;  T12.5-50A240;  T12.5-50B205;  T12.5-50B225;  T12.5-50B240;  T12.5-50C205;  T12.5-50C225;  T12.5-50C240;  T12.5-50D205;  T12.5-50D225;  T12.5-50D240;  T12.5-50E205;  T12.5-50E225;  T12.5-50E240;  T12.5-80A255;  T12.5-80A285;  T12.5-80A315;  T12.5-80A345;  T12.5-80A375;  T12.5-80B255;  T12.5-80B285;  T12.5-80B315;  T12.5-80B345;  T12.5-80B375;  T12.5-80C255;  T12.5-80C285;  T12.5-80C315;  T12.5-80C345;  T12.5-80C375;  T12.5-80D255;  T12.5-80D285;  T12.5-80D315;  T12.5-80D345;  T12.5-80D375;  T12.5-80E255;  T12.5-80E285;  T12.5-80E315;  T12.5-80E345;  T12.5-80E375;        trauma intramedullary rod, 8.5 mm diameter:   T8.5-50A205;  T8.5-50A225;  T8.5-50A240;  T8.5-50B205;  T8.5-50B225;  T8.5-50B240;  T8.5-50C205;  T8.5-50C225;  T8.5-50C240;  T8.5-50D205;  T8.5-50D225;  T8.5-50D240;  T8.5-50E205;  T8.5-50E225;  T8.5-50E240;  T8.5-80A255;  T8.5-80A285;  T8.5-80A315;  T8.5-80A345;  T8.5-80A375;  T8.5-80B255;  T8.5-80B285;  T8.5-80B315;  T8.5-80B345;  T8.5-80B375;  T8.5-80C255;  T8.5-80C285;  T8.5-80C315;  T8.5-80C345;  T8.5-80C375;  T8.5-80D255;  T8.5-80D285;  T8.5-80D315;  T8.5-80D345;  T8.5-80D375;  T8.5-80E255;  T8.5-80E285;  T8.5-80E315;  T8.5-80E345;  T8.5-80E375;    GTIN Number (covers both Humerus and Tibia):  00812258023836;  00812258023843;  00812258023850;  00812258023119;  00812258023126;  00812258023133;  00812258023591;  00812258023607;  00812258023614;  00812258023355;  00812258023362;  00812258023379;  00812258024079;  00812258024086;  00812258024093;  00812258023867;  00812258023874;  00812258023881;  00812258023898;  00812258023904;  00812258023140;  00812258023157;  00812258023164;  00812258023171;  00812258023188;  00812258023621;  00812258023638;  00812258023645;  00812258023652;  00812258023669;  00812258023386;  00812258023393;  00812258023409;  00812258023416;  00812258023423;  00812258024109;  00812258024116;  00812258024123;  00812258024130;  00812258024147;  00812258023911;  00812258023928;  00812258023935;  00812258023195;  00812258023201;  00812258023218;  00812258023676;  00812258023683;  00812258023690;  00812258023430;  00812258023447;  00812258023454;  00812258024154;  00812258024161;  00812258024178;  00812258023942;  00812258023959;  00812258023966;  00812258023973;  00812258023980;  00812258023225;  00812258023232;  00812258023249;  00812258023256;  00812258023263;  00812258023706;  00812258023713;  00812258023720;  00812258023737;  00812258023744;  00812258023461;  00812258023478;  00812258023485;  00812258023492;  00812258023508;  00812258024185;  00812258024192;  00812258024208;  00812258024215;  00812258024222;  00812258023751;  00812258023768;  00812258023775;  00812258023034;  00812258023041;  00812258023058;  00812258023515;  00812258023522;  00812258023539;  00812258023270;  00812258023287;  00812258023294;  00812258023997;  00812258024000;  00812258024017;  00812258023782;  00812258023799;  00812258023805;  00812258023812;  00812258023829;  00812258023065;  00812258023072;  00812258023089;  00812258023096;  00812258023102;  00812258023546;  00812258023553;  00812258023560;  00812258023577;  00812258023584;  00812258023300;  00812258023317;  00812258023324;  00812258023331;  00812258023348;  00812258024024;  00812258024031;  00812258024048;  00812258024055;  00812258024062;"
    }
  ]
}