      "status": "Terminated",
      "city": "Chesterfield",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84638",
      "recalling_firm": "ulrich medical USA Inc",
      "address_1": "18221 Edison Ave",
      "address_2": "N/A",
      "postal_code": "63005-3703",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AZ, TX, GA, FL, AR, CA",
      "recall_number": "Z-1350-2020",
      "product_description": "Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw",
      "product_quantity": "179 units",
      "reason_for_recall": "Due to a production error, the weld seam of the \"Locking screw for crosslink, screw-to-screw\" cannot reliably transmit the required torque of 3.5 Nm during installation.",
      "recall_initiation_date": "20191231",
      "center_classification_date": "20200224",
      "termination_date": "20251117",
      "report_date": "20200304",
      "code_info": "Lot Numbers:  U010577, U016620, U019084"
    }
  ]
}