  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84649",
      "recalling_firm": "Randox Laboratories, Limited",
      "address_1": "Ardmore; 55 The Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution including state of PR.",
      "recall_number": "Z-1349-2020",
      "product_description": "RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage:  A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.",
      "product_quantity": "U.S.: RX4040-2 units RX 4041-3 units",
      "reason_for_recall": "Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as \"out of range.\"",
      "recall_initiation_date": "20191219",
      "center_classification_date": "20200224",
      "termination_date": "20210423",
      "report_date": "20200304",
      "code_info": "Software version UI2550642107 for the RX Daytona Plus instrument."
    }
  ]
}