  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73429",
      "recalling_firm": "Ortho-Clinical Diagnostics",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "N/A",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US (nationwide) including Puerto Rico  and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.",
      "recall_number": "Z-1348-2016",
      "product_description": "VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110.  Intended for use in the in vitro measurement of a variety of analytes of clinical interest.",
      "product_quantity": "2,101 Total units (1,043 domestically & 1,058 internationally)",
      "reason_for_recall": "The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situation where a module that is intended to be de-energized will remain energized only if the associated AC power cord is unplugged.",
      "recall_initiation_date": "20160216",
      "center_classification_date": "20160407",
      "termination_date": "20180530",
      "report_date": "20160413",
      "code_info": "Product Codes: 6802413, 6802915 (Refurbished)  Serial numbers: J56000024 - J56002387",
      "more_code_info": ""
    }
  ]
}