  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73555",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of SC,  NC and the countries of Australia, Canada, France. Germany, Japan and Turkey.",
      "recall_number": "Z-1347-2016",
      "product_description": "SIGNA Pioneer MR Systems    Product Usage:  The SIGNA Pioneer is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the SIGNA Pioneer reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.",
      "product_quantity": "16 (2 US; 14 OUS)",
      "reason_for_recall": "Ferrous material in the  SIGNA Pioneer Table Control Box (TCB)  could cause unexpected attraction to the MR system's high strength  magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from  the magnetic field. There have been no events reported as a result of this issue. Clinical  scanni",
      "recall_initiation_date": "20160301",
      "center_classification_date": "20160407",
      "termination_date": "20160816",
      "report_date": "20160413",
      "code_info": "Mfg Lot or Serial #   System ID    001612TA1                N/A  001718TA6               336397P3T   002331TA7              083132425984  001719TA4             250598MR1  001424TA1             M2850013  001537TA0             TP0008MR07  001662TA6             157022MR01  001184TA1             ZC0002  001652TA7               N/A  002227TA7             ZC0003  002278TA0             ZC0004  002330TA9             ZC0006  002529TA6             N/A  001485TA2            34026MRS05  002279TA8            34225MRS02  002395TA2            34185MRS02",
      "more_code_info": ""
    }
  ]
}