  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87563",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of FL, MN, OH, PA, TX.",
      "recall_number": "Z-1346-2021",
      "product_description": "Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital  X-Ray System generate Xray images from the whole body including the skull, chest, abdomen, and extremities.",
      "product_quantity": "12 units",
      "reason_for_recall": "A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted",
      "recall_initiation_date": "20210310",
      "center_classification_date": "20210402",
      "termination_date": "20231211",
      "report_date": "20210414",
      "code_info": "Serial Numbers:  10041  10046  10055  10057  10053  10049  10043  10039  10054  10048  10042  10045"
    }
  ]
}