  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64930",
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "address_1": "1080 Us Highway 202 S",
      "address_2": "N/A",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-1346-2013",
      "product_description": "cobas¿ 4800 system \u0013z480 instrument.      Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.",
      "product_quantity": "425 instruments",
      "reason_for_recall": "In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system.  Raw data review indicates that, in these cases, all signals are shifted by one channel:  channel 1 results report as channel 2,  channel 2 results report as channel 3,  channel 3 results report as channel 4, and  channel 4 results report as channel 1.    In the reported",
      "recall_initiation_date": "20130401",
      "center_classification_date": "20130521",
      "termination_date": "20140807",
      "report_date": "20130529",
      "code_info": "Material Number 05200881001 - all lot/serial numbers."
    }
  ]
}