  "results": [
    {
      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72957",
      "recalling_firm": "AngioDynamics Inc.",
      "address_1": "10 Glens Falls Tech Park",
      "address_2": "N/A",
      "postal_code": "12801-3864",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide in the states of CA, IL, MA, SC, TX",
      "recall_number": "Z-1345-2016",
      "product_description": "NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941    Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire",
      "product_quantity": "8 kits",
      "reason_for_recall": "Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification.  Specifically, at least a portion of the affected lot is unable to accept a 0.038\" guidewire as intended.  In some cases the inner lumen was found to be 0.036\".",
      "recall_initiation_date": "20151103",
      "center_classification_date": "20160407",
      "termination_date": "20170316",
      "report_date": "20160413",
      "code_info": "Batch/Lo. 4925662 Use By 2017-02",