  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90407",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL,  IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria,  Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy,  Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.",
      "recall_number": "Z-1343-2022",
      "product_description": "Artis zeego (Model no.  10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System",
      "product_quantity": "233 worldwide; 32 US",
      "reason_for_recall": "It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.",
      "recall_initiation_date": "20220524",
      "center_classification_date": "20220707",
      "termination_date": "20250122",
      "report_date": "20220713",
      "code_info": "UDI-DI: Artis zeego 04056869010083 Artis Q.zeego 04056869010007",
      "more_code_info": ""
    }
  ]
}