  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72957",
      "recalling_firm": "AngioDynamics Inc.",
      "address_1": "10 Glens Falls Tech Park",
      "address_2": "N/A",
      "postal_code": "12801-3864",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide in the states of CA, IL, MA, SC, TX",
      "recall_number": "Z-1343-2016",
      "product_description": "NAMIC CONVENIENCE KIT SELF REGIONAL HEALTHCARE PV LEG PACK, REF Catalog No. 651919311, UPN Product No. H7496519193111    Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire",
      "product_quantity": "136 kits",
      "reason_for_recall": "Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification.  Specifically, at least a portion of the affected lot is unable to accept a 0.038\" guidewire as intended.  In some cases the inner lumen was found to be 0.036\".",
      "recall_initiation_date": "20151103",
      "center_classification_date": "20160407",
      "termination_date": "20170316",
      "report_date": "20160413",
      "code_info": "Batch/Lot No. 4921353 Use By 2018-05 and Batch/Lot No. 4932710 Use By 2018-06-30",
      "more_code_info": ""
    }
  ]
}